FDA Device Recalls

Recalls /

#102284

Product

Smith & Nephew, Tendon Stripper, Slotted, 5.0 mm I.D. x 12 working length. Non-Sterile Catalog Number: 013554 Orthopedic Manual Surgical Instrument

FDA product code
LXHOrthopedic Manual Surgical Instrument
Device class
Class 1
Medical specialty
Orthopedic
Affected lot / code info
Lot Numbers: 50342499, 50344799, 50347017, 50349339 50354170, 50357485, 50360691, 50360992 50367247, 50374544, 50375820, 50370949 50371944

Why it was recalled

Shaft tip (casting) does not meet the SNE 416 specification, which could result in rusting of the Tendon Strippers

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Smith & Nephew sent an Urgent - Product Recall Notification letter dated July 7, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete the form included in the letter and return to Hoangthi.le@smith-nephew.com or fax to 508-261-3620. Each affiliate must then notify each facility affected by the recall. To return product and obtain credit or order replacement product customers should contact the Smith & Nephew Endoscopy Returns at 800-343-5717 (option 3) for a Return Authorization (RA) or send an email to, endo.andreturns@smith-nephew.com. For questions, contact Hoangthi by phone at 508-337-3731.

Recalling firm

Firm
Smith & Nephew, Inc. Endoscopy Division
Address
150 Minuteman Drive, Andover, Massachusetts 01810-1031

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) and the countries of Australia, Austria, Belgium, Canada, Chile, China, Columbia, Denmark, Dubai, Finland, France, Germany, Greece, Hong Kong, India, Italy, Israel, Japan, Korea, Malaysia, Mexico, Netherland, Norway, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, and Venezuela

Timeline

Recall initiated
2011-07-07
Posted by FDA
2011-08-22
Terminated
2015-07-23
Status

Source: openFDA Device Recall endpoint. Recall record ID #102284. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.