Product

Dimension Calcium Flex Reagent is used on the Dimension clinical chemistry system as an in vitro diagnostic test intended for the quantitive determination of calcium activity in human serum, plasma and urine. Dimension(R) CA Flex(R) reagent is used on the Dimension(R) clinical chemistry system as an in vitro diagnostic test intended for the quantitative determination of calcium activity in human serum, plasma and urine.

FDA product code
JJEAnalyzer, Chemistry (Photometric, Discrete), For Clinical Use
Device class
Class 1
Medical specialty
Clinical Chemistry
510(k) numbers
K860021
Affected lot / code info
Catalog # DF23A; Lot # EA2153

Why it was recalled

Siemens Healthcare Diagnostic has confirmed a high frequency of "Abnormal assay" flags and imprecision. Test results across the assay range may be falsely depressed or elevated, as much as, 1.93 mg/dL at a concentration of 9.26 mg/dL.

Root cause (FDA determination)

Other

Action the firm took

An Urgent Field Safety Notice datey July 2011 was issued to all affected customers. Customers are being advised to immediately discontinue use and discard any remaining inventory of Dimension(R) CA Flex(R0 reagent cartridge - lot EA2153. Replacement order for the affected lot is in process, and will be sent to impacted customers free of charge. Customers were instructed to forward the Field Safety Notice to anyone whom they may have distributed the affected product. For questions regarding this recall call 302-631-6299.

Recalling firm

Firm
Siemens Healthcare Diagnostics, Inc.
Address
500 GBC Drive PO BOX 6101, Mailstop 514, Newark, Delaware 19714-6101

Distribution

Distribution pattern
Worldwide Distribution to Guam, Puerto Rico, Chile, Costa Rica, Venezuela, South Korea, Vietnam and domestically to AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MN, MI, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WV, WI, and WY.

Timeline

Recall initiated
2011-07-19
Posted by FDA
2011-08-19
Terminated
2013-12-27
Status

Source: openFDA Device Recall endpoint. Recall record ID #102582. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.