FDA Device Recalls

Recalls /

#103986

Product

SARNS INTRACARDIAC SUCKER, BULK, NON-STERILE

FDA product code
DTSSucker, Cardiotomy Return, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
Affected lot / code info
Catalog number 11009010

Why it was recalled

Terumo Cardiovascular Systems (Terumo CVS) recently discovered three instances of metal fragments inside the stainless steel tube of the Sarns" Adult Rigid Intracardiac Sucker.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

On 9/13/2011 Terumo sent a letter entitled "URGENT MEDICAL DEVICE REMOVAL" to its customers that included the name and description of the device, lot numbers, the potential hazard, correction, the reason for the recall and instructions. Customers are to review the notice, ensure that all users are aware of the issue, and a recall response form to be completed and returned. Customers are to call Terumo CVS Customer Service to obtain a Return Goods Authorization number. The letter also included MedWatch reporting information and stated that Customers with questions should call 1-800-521-2818.

Recalling firm

Firm
Terumo Cardiovascular Systems Corporation
Address
6200 Jackson Road, Ann Arbor, Michigan 48103-9586

Distribution

Distribution pattern
Worldwide Distribution -- USA, Belgium, Australia, Mexico, and Japan.

Timeline

Recall initiated
2011-09-13
Posted by FDA
2011-10-17
Terminated
2012-12-10
Status

Source: openFDA Device Recall endpoint. Recall record ID #103986. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.