Plum A+ Hyperbaric Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; list 11005. Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged the clamp prior to opening the cassette door, unrestricted flow may occur.
- FDA product code
- FRN — Pump, Infusion
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K081412
- Affected lot / code info
- list 11005, serial numbers 0016792207 through 0018807700
Why it was recalled
Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged the clamp prior to opening the cassette door, unrestricted flow may occur.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
Hospira, Inc. sent an "URGENT DEVICE RECALL" letter dated September 16, 2011 to all affected customers. The letter describes the product, problem and actions to be taken by the customers. The letter instructs customers on how to prevent occurrences of continuous recycling/rebooting during start-up. A reply form was attached to the letter for customers to complete and return via fax to 1-888-345-5358. Contact Hospira Advanced Knowledge Center at 1-800-241-4002, option 4, for technical assistance or information regarding this notice.
Recalling firm
- Firm
- Hospira Inc.
- Address
- 275 N Field Dr, Lake Forest, Illinois 60045-2579
Distribution
- Distribution pattern
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the U. S. Virgin Islands and countries of Argentina, Australia, Bahamas, Barbados, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United kingdom, United Arab Emirates, Uruguay and Vietnam.
Timeline
- Recall initiated
- 2011-09-16
- Posted by FDA
- 2011-09-29
- Terminated
- 2016-11-07
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #104074. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.