FDA Device Recalls

Recalls /

#104075

Product

Plum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971 b) list 11973 - software version 10.3 c) list 12391 - software version 11.3 d) list 20679 - Hospira MedNet Software e) list 20792 - Driver Product Usage: The Plum A+ dual-line volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With their primary line, secondary line, and piggyback fluid delivery capability, the Plum A+ is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories and the LifeShield needleless protection systems makes the Plum A+ a convenient and cost-effective infusion system.

FDA product code
FRNPump, Infusion
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K011442, K042081, K070398, K982159
Affected lot / code info
a) list 11971- serial numbers 0013120799 through 0099072300; b) list 11973 - serial numbers 0012570058 through 0012573663; c) list 12391 - serial numbers 0000073363 through 0999948294; d) list 20679 - serial numbers 0015441179 through 0015459047; e) list 20792 - serial numbers 0016061555 through 0026347395

Why it was recalled

Hospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This condition could cause a delay in therapy if the clinician is unable to start the device due to the continuous recycling and/or rebooting.

Root cause (FDA determination)

Software design

Action the firm took

The firm, Hospira sent "URGENT DEVICE RECALL" letters dated September 16, 2011 to all affected customers. The letters identified the products, problem and action(s) to be taken. The customers were instructed to read the information provided carefully, complete the attached Reply Form and return it to Hospira via fax at 1-888-345-5358. The letter states that a software upgrade to address the issue is currently being developed. Customers were instructed to notify their consignees if they have further distributed the affected product. The customers were requested to contact Hospira Advanced Knowledge Center at 1-800-241-4002, option 4, for technical assistance. *** Follow-up letters dated June 26, 2012 were sent to the accounts listing the device modules with software versions v13.2, v13.3, v13.40 and v13.5 that will require the software upgrade. ***

Recalling firm

Firm
Hospira Inc.
Address
275 N Field Dr, Lake Forest, Illinois 60045-2579

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) including Puerto Rico and the US Virgin Islands, and countries of: Argentina, Australia, Bahamas, Barbados, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United kingdom, United Arab Emirates, Uruguay and Vietnam.

Timeline

Recall initiated
2011-09-16
Posted by FDA
2011-10-04
Terminated
2016-11-07
Status

Source: openFDA Device Recall endpoint. Recall record ID #104075. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.