Product

cobas c 311 Analyzer, Part Number: 04826876001 In-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.

FDA product code
JJEAnalyzer, Chemistry (Photometric, Discrete), For Clinical Use
Device class
Class 1
Medical specialty
Clinical Chemistry
Affected lot / code info
Serial Numbers:1050-16, 1040-08, 1041-06, 0927-16, 0918-05, 1048-02, 0812-12, 1058-09, 0925-20, 1044-19, 1041-05, 1043-15, 1056-18, 0812-09, 1038-07, 0814-10, 1041-19, 1170-04, 0929-08, 0919-16, 0814-03, 0926-11, 1035-01, 1034-19, 1035-10, 1174-19, 1176-03, 1067-02, 0919-17, 0920-05, 0927-13, 1060-17, 1060-05, 0927-17, 1044-18, 0812-10, 0811-20, 1041-20, 1174-04, 0925-05, 1060-01, 1062-09, 1066-20, 1174-07, 0814-15, 0814-17, 1053-16, 1054-02, 1034-02, 1052-16, 1052-19, 0918-06, 1052-13, 1052-20, 1052-15, 1034-07, 1043-10, 0811-15, 1054-03, 1061-08, 1043-06, 0926-10, 1044-06, 1034-03, 1052-09, 1043-12, 1055-01, 0929-09, 1038-05, 1034-01, 0810-19, 1054-20, 1062-07, 1059-20, 0814-16, 1043-14, 1062-08, 1054-01, 0929-07, 1038-06, 1056-19, 1057-03, 1169-20, 0926-19, 1067-01, 1038-04, 0925-06, 1043-08, 1043-13, 0812-11, 1057-05, 1060-06, 0926-16, and 1034-06.

Why it was recalled

The cobas c 311 analyzer software may incorrectly calculate the volume of the ISE Reference Electrolyte solution remaining on the analyzer. In some cases the incorrect results obtained will not be flagged with any data alarms.

Root cause (FDA determination)

Software design

Action the firm took

October 26, 2011 Urgent Medical Device Corrections were sent to all direct accounts in the USA that have installed the cobas c 311 analyzer and included a faxback form to be completed and faxed to 1-877-766-7452. Consignees were instructed to temporarily replace the ISE Reference Electrolyte bottle based upon the frequency calculated from the maximum number of ISE samples run per day and activate the yellow threshold alarm on the system. Customers with question and concerns are directed to call Roche Diagnostics Technical Support at 1-800-428-2366, available 24-hours, 7 days a week.

Recalling firm

Firm
Roche Diagnostics Operations, Inc.
Address
9115 Hague Road, Indianapolis, Indiana 46256-1025

Distribution

Distribution pattern
Worldwide Distribution -- USA including the states of Maryland and Texas.

Timeline

Recall initiated
2011-10-26
Posted by FDA
2011-11-30
Terminated
2012-07-09
Status

Source: openFDA Device Recall endpoint. Recall record ID #104939. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: cobas c 311 Analyzer, Part Number: 04826876001 In-vitro diagnostic analyzer capable of performing c · FDA Device Recalls