Recalls / —
—#104990
Product
Stryker Endoscopy 3.55 MM Super 90 S-SERFAS Energy Probe; electrosurgical device. Model number: 279-351-300;
- FDA product code
- GEI — Electrosurgical, Cutting & Coagulation & Accessories
- Device class
- Class 2
- Medical specialty
- General, Plastic Surgery
- 510(k) numbers
- K071859
- Affected lot / code info
- Model number: 279-351-300; Lots 11161AE2 to 11241AE2, Lots are numbered sequentially using the middle three numbers, i.e 161-241.
Why it was recalled
An increased number of complaints for tip breaks were observed for the 3.5mm Super 90-S SERFAS Energy Suction Probes manufactured from June to August 2011.
Root cause (FDA determination)
Component design/selection
Action the firm took
Stryker Endoscopy Urgent Medical Device Removal notifications datyed October 25, 2011, were sent to all consignees via traceable mail. Sales representatives and regional managers were notified by e-mail. Product is to be returned to Stryker Endoscopy using the pre-paid shipping label. Replacement product may be obtained by calling Stryker at 1-800-624-4422. The enclosed Acknowledgement receipt form is to be completed and returned. Questions or comments may be directed to Stryker at 1-800-624-4422 or e-mail at 90S@stryker.com.
Recalling firm
- Firm
- Stryker Endoscopy
- Address
- 5900 Optical Ct, San Jose, California 95138-1400
Distribution
- Distribution pattern
- Worldwide distribution including USA, Australia, Canada, China, Greece, Hong Kong, India, Italy, Japan, France, Netherlands, Poland, South Africa, Spain, Sweden, Switzerland and the United Kingdom
Timeline
- Recall initiated
- 2011-10-24
- Posted by FDA
- 2011-11-21
- Terminated
- 2012-02-29