FDA Device Recalls

Recalls /

#104990

Product

Stryker Endoscopy 3.55 MM Super 90 S-SERFAS Energy Probe; electrosurgical device. Model number: 279-351-300;

FDA product code
GEIElectrosurgical, Cutting & Coagulation & Accessories
Device class
Class 2
Medical specialty
General, Plastic Surgery
510(k) numbers
K071859
Affected lot / code info
Model number: 279-351-300; Lots 11161AE2 to 11241AE2, Lots are numbered sequentially using the middle three numbers, i.e 161-241.

Why it was recalled

An increased number of complaints for tip breaks were observed for the 3.5mm Super 90-S SERFAS Energy Suction Probes manufactured from June to August 2011.

Root cause (FDA determination)

Component design/selection

Action the firm took

Stryker Endoscopy Urgent Medical Device Removal notifications datyed October 25, 2011, were sent to all consignees via traceable mail. Sales representatives and regional managers were notified by e-mail. Product is to be returned to Stryker Endoscopy using the pre-paid shipping label. Replacement product may be obtained by calling Stryker at 1-800-624-4422. The enclosed Acknowledgement receipt form is to be completed and returned. Questions or comments may be directed to Stryker at 1-800-624-4422 or e-mail at 90S@stryker.com.

Recalling firm

Firm
Stryker Endoscopy
Address
5900 Optical Ct, San Jose, California 95138-1400

Distribution

Distribution pattern
Worldwide distribution including USA, Australia, Canada, China, Greece, Hong Kong, India, Italy, Japan, France, Netherlands, Poland, South Africa, Spain, Sweden, Switzerland and the United Kingdom

Timeline

Recall initiated
2011-10-24
Posted by FDA
2011-11-21
Terminated
2012-02-29
Status

Source: openFDA Device Recall endpoint. Recall record ID #104990. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.