Recalls / —
—#105471
Product
Roller Pump, 4 Inch Diameter The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
- FDA product code
- DTQ — Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K022947
- Affected lot / code info
- Catalog Number 816570; Serial Numbers: 5939 - 5962
Why it was recalled
Terumo Cardiovascular Systems (Terumo CVS) discovered and rejected a single digital encoder cable assembly during incoming inspection because it had a loose wire. The assembly is used in the 4" and the 6" roller pumps for the Terumo Advanced Perfusion System 1.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
TERUMO sent an Urgent Medical Device Correction letter dated January 13, 2012, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to review the Medical Device Correction notice. Assure that all users are aware of the notice. Confirm receipt of the communication by faxing the attached Customer Response Form to 1-900-292-6551. Roller pumps with affected cables will have the cables assembly replaced. For questions regarding this recall call 1-800-521-2818.
Recalling firm
- Firm
- Terumo Cardiovascular Systems Corporation
- Address
- 6200 Jackson Road, Ann Arbor, Michigan 48103-9586
Distribution
- Distribution pattern
- Worldwide Distribution - USA including MD, WI, and IA and the countries of Canada, Mexico, and Vietnam
Timeline
- Recall initiated
- 2012-01-13
- Posted by FDA
- 2012-01-24
- Terminated
- 2012-12-03
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #105471. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.