Product
HOTLlNE Disposable Administration Sets, Product Code L-370 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE Routine Flow Blood and Fluid Warmers (HL-90 and HL-290), for the warming and delivery of blood and I.V. solutions, at normothermic flow rates of 50 - 5,000 ml/hr.
- FDA product code
- LGZ — Warmer, Thermal, Infusion Fluid
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K911383
- Affected lot / code info
- Lot Numbers: 1995695 1995696 2038150 2038151 2038152 2047423 2061923 2061924 2061929 2061930 2061931 2061932 2061933 2077201 2077202 2090780 2090781 2095295 2095302 2095306
Why it was recalled
Disconnected/ loose Luer connectors at the patient end of the triple lumen tubing of the HOTLlNE Disposable Administration Sets and Level 1 D/DI-60HL Normothermic IV Administration Sets
Root cause (FDA determination)
Device Design
Action the firm took
Smiths Medical notified consignees by letter dated November 28, 2011 titled: Urgent Field Safety Notice. The letter identified the affected product, the problem, and the actions to be taken. Customers were instructed to inspect inventory and segregate unused affected product. Distributors were instructed to immediately stop distributing and quarantine all inventory. The letters included a Confirmation Form to be complete and return, by fax to 781-610-9859 or by email to hotlineconnectorrecall@smiths-medical.com. For questions consignees were instructed to contact Smiths Medical at 1781-763-9330.
Recalling firm
- Firm
- Smiths Medical ASD, Inc.
- Address
- 160 Weymouth St, Rockland, Massachusetts 02370-1136
Distribution
- Distribution pattern
- Worldwide Distribution (USA) (nationwide) - Argentina, Austria, Australia, Belgium, Bahrain, Bermuda, Canada, Chile, China, Columbia, Czech Republic, Germany, Denmark, Estonia, Egypt, Finland, France, Great Britain, Guernsey, Greece, Guatemala, Hong Kong, Hungary, Indonesia, Ireland, Israel India, Iceland, Italy, Japan, South Korea, Kazakhstan, Lebanon, Luxembourg, Mexico, Malaysia, New Caledonia, Netherlands, Norway, New Zealand, Panama, Poland, Portugal, Qatar, Singapore, Sweden, Slovak Republic, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, and Vatican City State.
Timeline
- Recall initiated
- 2011-11-21
- Posted by FDA
- 2012-01-11
- Terminated
- 2014-08-07
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #105684. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.