Product

Biomet Series A Thin Patella size 28mm, three pegs - 184782, Sterile, Biomet Orthopedics, Warsaw, IN 46582 For use with Vanguard, Maxim, Ascent & AGC Knee Systems Knee joint patellofemorotibial, cemented, polymer/metal/polymer, semi-constrained. Knee joint replacement prosthesis, intended for patella replacement of the knee joint in conjunction with a femoral and tibial component.

FDA product code
JWHProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K040770
Affected lot / code info
Part 184782: Lot 098970, 144970, 190440, 232540, 274550, 357320, 397810, 397820, 438790, 438800, 498310, 539710, 599220, 644430, 710240

Why it was recalled

Correcting language on product labeling which incorrectly states patella buttons are intended for use with the AGC knee system.

Root cause (FDA determination)

Error in labeling

Action the firm took

Biomet sent a "FIELD CORRECTION NOTICE" dated October 3, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to locate the affected product and to follow the work instructions included in the letter. Additionally, a Confirmation of Completion Response Form was enclosed for customers to complete and return via fax to 574-372-1683. Questions related to this notice are directed to 574-371-3755.

Recalling firm

Firm
Biomet, Inc.
Address
56 E Bell Dr, Warsaw, Indiana 46582-6989

Distribution

Distribution pattern
Worldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Argentina, Australia, Canada, Chile, Costa Rica, Europe, India, Japan, Korea, and Singapore.

Timeline

Recall initiated
2011-10-04
Posted by FDA
2012-01-11
Terminated
2013-04-08
Status

Source: openFDA Device Recall endpoint. Recall record ID #105833. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: Biomet Series A Thin Patella size 28mm, three pegs - 184782, Sterile, Biomet Orthopedics, Warsaw, IN · FDA Device Recalls