Product

Baxter Interlink System Continu-Flo Solution Set with Control-A-Flo Regulator, 89" (2.3 m), 2 Injection Sites, Male Luer Lock Adapter; an Rx sterile, nonpyrogenic fluid pathway for the administration of sterile I.V. fluids to the patient; Baxter Healthcare Corporation, Deerfield, IL 60015 USA; product code 1C8621 Intended use: for the administration of drug and solutions to patient's vascular system through a vascular access device

FDA product code
FPASet, Administration, Intravascular
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K940867
Affected lot / code info
product code 1C8621, lot numbers GR295352, GR296095 and GR296335

Why it was recalled

During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection.

Root cause (FDA determination)

Employee error

Action the firm took

The firm, Baxter, sent an "Urgent Product Recall" letter dated December 6, 2011, via first class mail to its customers. The letter described the product, problem, and actions to be taken. The customers were instructed to inspect all locations where their facility's inventory is located for products with the affected lot numbers; return the affected product; request a credit or replacement by calling Baxter Healthcare Center for Service at 1-888-229-0001 between 7:00am to 6:00pm Central time, Monday through Friday; complete and return the attached Customer Reply Form via fax to: 1-847-270-5457, or scan and email to fca@baxter.com; and if they are a dealer, wholesaler, or distributor/reseller that distributed any affected product to other facilities, please notify their customers of this recall. If you have any questions regarding the communication, please call The Center for One Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CST.

Recalling firm

Firm
Baxter Healthcare Corp. Rt.
Address
120 & Wilson Rd, Round Lake, Illinois 60073

Distribution

Distribution pattern
Worldwide distribution: USA (nationwide) including states of: Alaska, California, Colorado, Florida, Georgia, Hawaii, Illinois, Kansas, Kentucky, Massachusetts, Maine, Michigan, Minnesota, Missouri, North Carolina, New Hampshire, New Jersey, New York, Ohio, Pennsylvania, Rhode Island, South Carolina, Tennessee, Virginia, Washington, Wisconsin and West Virginia; and country of: New Zealand.

Timeline

Recall initiated
2011-12-06
Posted by FDA
2012-01-11
Terminated
2012-11-01
Status