Siemens MAGNETOM Aera/Skyra, model numbers 10432914 and 10432915. Nuclear magnetic resonance imaging system.

FDA product code
LNHSystem, Nuclear Magnetic Resonance Imaging
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K101347
Affected lot / code info
Serial numbers 45115, 41187, 41157, 41116, 45072, 45081, 41089, 41215, 41144, 41201, 45148, 41164, 41088, 41245, 41168, 41152, 45140, 45152, 45143, 45195, 41222, 41150, 45129, 45118, 45119, 41134, 41138, 45182, 45213, 41167, 41100, 52024, 45196, 45197, 41180, 41130, 45088, 41189, 41104, 41246, 41176, 45138, 15084, 41071, 45120, 45107, 41271, 45135, 45141, 41264, 45110, 41279, 41231, 41284, 41235, 45221, 45223, 41276, 41278, 41137, 41225, 41131, 41141, 41233, 45185, 41236, 41238, 45149, 41191, 45074, 45106, 45006, 45201, 41133, 41265, 41188, 45220, 41140, 41079, 45160, 45167, 41194, 45155, 45130, 41241, 41170, 45147, 41006, 45219, 45085, 45150, 45031, 41162, 45121, 45134, 45099, 45038, 45079, 41109, 41156, 45116, 41249, 45128, 45078, 45214, 41103, 45229, 41234, 45181, 52025, 52021, 45105, 45132, 41149, 45228, 45060, 45080, 45208, 41083, 45230, 45205, 41097, 41181, 45093, 45199, 45184, 45071, and 45111.

Why it was recalled

During product monitoring firm has discovered that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing necessary. The potential existed for a cable connection to become overheated, leading to the emission of smoke and possible material damage. Firm preventively checked the gradient cabl

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Siemens Medical Solutions sent a "CUSTOMER SAFETY ADVISORY NOTICE" letter dated November 24, 2010 to all affected customers. The letter described the product, problem, and actions to be taken. Contact the firm at 610-219-4834 for questions regarding this recall.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
51 Valley Stream Pkwy, Malvern, Pennsylvania 19355

Distribution

Distribution pattern
Nationwide Distribution-including the states of AK, AZ, CA, CT, DE, FL, GA, IL, IN, KS, MD, MA, MI, MN, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, and WI.

Timeline

Recall initiated
2012-01-26
Posted by FDA
2012-02-02
Terminated
2014-01-31
Status

Source: openFDA Device Recall endpoint. Recall record ID #107139. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.