Recalls / —
—#108392
Product
Siemens-branded linear accelerator with a 550 TxT Treatment Table with a serial number <_ 1894. Siemens brand ARTISTE, ONCOR and PRIMUS Linear Accelerator systems include the 550 TxT" Treatment Table. Manufactured by: SIEMENS AG, MEDICAL SOLUTIONS Roentgenstrasse 19-21 Kemnath, GERMANY 95478. Product Usage: The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
- FDA product code
- IYE — Accelerator, Linear, Medical
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K060226, K103606
- Affected lot / code info
- System IVK no. Component Component IVK Ser # ARTISTE MV system 8139789 550 TxT table 7346534 d1894 ONCOR Expression 7360717 550 TxT table 7346534 d1894 ONCOR Impression 5857912 550 TxT table 7346534 d1894 Plus ONCOR Impression 5857920 550 TxT table 7346534 d1894 ONCOR Avant Garde 5863472 550 TxT table 7346534 d1894 PRIMUS HI 4504200 550 TxT table 7346534 d1894 PRIMUS Mid-Energy 1940035 550 TxT table 7346534 d1894
Why it was recalled
SIEMENS has received reports where a dropping down of the 550 TxT" Treatment Table was observed while the table was being moved downwards during operation by means of the hand control.
Root cause (FDA determination)
Software design
Action the firm took
Siemens sent Customer Information letters to all affected customers beginning April 3, 2012 via Siemens Field Service Engineers. The letter identified the affected product, what is the issue and what has Siemens done to address this issue. The letter included an Update Instructions TH006/12/S - TxT Table Software Update VD00A. SIEMENS has arranged for a modification of the affected system. A SIEMENS Service Representative will ensure that the new software version VD00A is installed on affected system. The letter instructs customer to please include this Customer Information in their Digital Linear Accelerator System Owner Manual chapter Safety Advisory Letters where it should remain.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 4040 Nelson Ave, Concord, California 94520-1200
Distribution
- Distribution pattern
- Worldwide Distribution - USA (nationwide) including P.R and the countries of: Africa, Angola, Argentina, Australia, Austria, Belgium, Bosnia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Egypt, France, Germany, Greece, Hungary, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Norway, Pakistan, Philippines, Poland, Portugal, Republic of Korea, Russian Federation, Saudi Arabia, Serbia, South Thailand, Sri Lanka, Spain, Switzerland, Trinidad, Tobago, Turkey, Ukraine and United Kingdom, .
Timeline
- Recall initiated
- 2012-04-03
- Posted by FDA
- 2012-05-03
- Terminated
- 2013-03-14
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #108392. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.