FDA Device Recalls

Recalls /

#108410

Product

GE Healthcare Automatic Mobile X- Ray (AMX) series: Brivo XR285 15KW, Brivo XR285 30KW, Optima XR200 15KW, Optima XR200 30KW, Optima XR220 15KW, and Optima XR220 30KW. General Electric, Waukesha, WI Indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.

FDA product code
IZLSystem, X-Ray, Mobile
Device class
Class 2
Medical specialty
Radiology
Affected lot / code info
All associated serial numbers.

Why it was recalled

GE Healthcare has become aware of a discrepancy on the x-ray source assembly of the mobile radiographic product, related to the light field to x-ray field edge alignment.

Root cause (FDA determination)

Radiation Control for Health and Safety Act

Action the firm took

GE Healthcare will issue an Important Electronic Product Radiation Warning letter to customers. The letter will describe the defect and related hazards, along with the safety instructions, affected product, and product correction GE Healthcare will, without charge, remedy the defect or bring the product into compliance with each applicable Federal and IEC standard in accordance with an approved plan, the details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit. A GE Healthcare Service Representative will update the external tube port on the source assembly to address the issue. Questions or concerns can be directed to the Call Center at 800-437-1171 or your local GE Healthcare Service Representative.

Recalling firm

Firm
GE Healthcare, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
USA (nationwide)

Timeline

Recall initiated
2011-12-23
Posted by FDA
2012-04-24
Terminated
2014-01-16
Status

Source: openFDA Device Recall endpoint. Recall record ID #108410. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: GE Healthcare Automatic Mobile X- Ray (AMX) series: Brivo XR285 15KW, Brivo XR285 30KW, Optima XR200 · FDA Device Recalls