Product

Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon Stationary Fluoroscopic system

FDA product code
MQBSolid State X-Ray Imager (Flat Panel/Digital Imager)
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K982795
Affected lot / code info
Software release 2.0.2 and 2.0.2SP1 including mirror icon

Why it was recalled

With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "right patient side" marker, although this can appear on the left patient side

Root cause (FDA determination)

Software change control

Action the firm took

Philips Healthcare issued a Field Safety Notice on 5/22/12 advising users of the error. A software update (Rel 2.1.2) will issue to correct the problem. Contact your local Philips representative. 1-800-722-9377. For questions regarding this recall call 978-687-1501.

Recalling firm

Firm
Philips Healthcare Inc.
Address
3000 Minuteman Road, Andover, Massachusetts 01810

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) and the countries of: Nationwide Foreign: Canada AFGHANISTAN ALBANIA ALGERIA ANGOLA ARGENTINA ARMENIA ARUBA AUSTRALIA AUSTRIA AZERBAIJAN BAHAMAS BAHRAIN BANGLADESH BARBADOS BELARUS BELGIUM BENIN BHUTAN BOLIVIA BOSNIA&HERZEGOVINA BRAZIL BRUNEI DARUSSALAM BULGARIA BURUNDI CAMBODIA CAMEROON CANARY ISLANDS CHAD CHILE CHINA COLOMBIA CONGO CONGO,DEMOCRATIC COSTA RICA COTE D'IVOIRE CROATIA CUBA CYPRUS CZECH REPUBLIC DENMARK DJIBOUTI DOMINICAN REPUBLIC ECUADOR EGYPT EL SALVADOR EQUATORIAL GUINEA ERITREA ESTONIA ETHIOPIA fake test ssd FINLAND FRANCE FRENCH GUIANA GABON GAMBIA GEORGIE GERMANY GHANA GREECE GUADELOUPE GUAM GUATEMALA GUINEA GUYANA HONDURAS HONG KONG HUNGARY ICELAND INDIA INDONESIA IRAN IRAQ IRELAND ISRAEL ITALY JAMAICA JAPAN JORDAN KAZAKSTAN KENYA KUWAIT KYRGYZSTAN LAO PEOPLE'S DEM.REP LATVIA LEBANON LIBERIA LIBYAN ARAB JAMAHIRI LIECHTENSTEIN LITHUANIA LUXEMBOURG MACAU MACEDONIA MALAWI MALAYSIA MALI MALTA MARTINIQUE MAURITANIA MAURITIUS METRACOM MEXICO MOLDOVA MONGOLIA MOROCCO MOZAMBIQUE NAMIBIA NEPAL NETHERLANDS NETHERLANDS ANTILLES NEW CALEDONIA NEW ZEALAND NICARAGUA NIGERIA NORTH KOREA NORWAY OMAN PAKISTAN PALESTINE PANAMA PARAGUAY PERU PHC BEST PHILIPPINES PMCC BU Canada PMCC BU USA POLAND PORTUGAL PUERTO RICO QATAR REUNION ROMANIA RUSSIA RWANDA SAUDI ARABIA SENEGAL SERBIA&MONTE NEGRO SIERRA LEONE SINGAPORE SLOVAKIA SLOVENIA SOMALIA SOUTH AFRICA SOUTH KOREA SPAIN SRI LANKA SUDAN SURINAME SWEDEN SWITZERLAND SYRIAN ARAB REPUBLIC TAIWAN TAJIKISTAN TANZANIA THAILAND TRINIDAD AND TOBAGO TUNISIA TURKEY TURKMENISTAN UGANDA UKRAINE Unidentified Country UNITED ARAB EMIRATES UNITED KINGDOM URUGUAY UZBEKISTAN VENEZUELA VIET NAM YEMEN ZAMBIA ZIMBABWE

Timeline

Recall initiated
2012-05-22
Posted by FDA
2012-06-15
Terminated
2017-02-24
Status

Source: openFDA Device Recall endpoint. Recall record ID #109499. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.