"***Electric Systems Foot Control with Direction and Irrigation***Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium.
- FDA product code
- HBE — Drills, Burrs, Trephines & Accessories (Simple, Powered)
- Device class
- Class 2
- Medical specialty
- Neurology
- 510(k) numbers
- K011444, K080802
- Affected lot / code info
- Serial Numbers: E3230327501, E32305327502, E32305327503, E32305327504, E32305327505, E32305327601, E32305327602, E32305327603, E32305327604, and E32305327605
Why it was recalled
The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
Anspach sent an Urgent Product Removal letter dated January 9, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to contact their sales representative to facilitate the removal action by providing a replacement for the affected product. For question contact your local sales representative or the Product Support department.
Recalling firm
- Firm
- The Anspach Effort, Inc.
- Address
- 4500 Riverside Drive, Palm Beach Gardens, Florida 33410-4235
Distribution
- Distribution pattern
- Worldwide Distribution - US (nationwide) and the countries of: Italy, Japan, Switzerland, Turkey and Taiwan.
Timeline
- Recall initiated
- 2012-01-09
- Posted by FDA
- 2012-10-18
- Terminated
- 2014-09-16