FDA Device Recalls

Recalls /

#114016

Product

bioMerieux Kit, Instrument & Update, B.40 Dualt-ML, Industry instrument software update, English, Spanish, Italian, German, French, Japanese, Brazilian, Portuguese, catalog number 412950 Software update for an instrument which is most commonly used to isolate and recover specimens drawn from adult and older pediatric patients that are acutely ill with signs and symptoms suggestive of blood stream infection, which can be due to a variety of sources including lung, abdomen, urinary tract, endovascular, and skin.

FDA product code
MDBSystem, Blood Culturing
Device class
Class 1
Medical specialty
Microbiology
Affected lot / code info
Identified by B.40 firmware

Why it was recalled

The product may mislabel patient information on culture specimen bottles due to a software miscommunication.

Root cause (FDA determination)

Software design

Action the firm took

BIOMERIEUX sent an Urgent Product Correction Notice to all affected customers starting on October 16, 2012, The letter identified the affected product, the problem, and the action to be taken by hte customer. The firm provided modified instructions for use to clarify how to properly load bottles and to provide appropriate cautions and warnings concerning bottle loading. Customers were instructed to ensure that the letter and guidance on how to properly scan and load bottles is distributed, and training is provided to all appropriate personnel within their organization. Check for lost bottles in their system; fill out and return the Acknowledgement Form in Attachment A by fax to confirm receipt of the notice. Customers were instructed to contact their local bioMerieux Customers Service Representative with any questions. For questions regarding this recall call 919-620-2682.

Recalling firm

Firm
Biomerieux Inc
Address
595 Anglum Rd, Hazelwood, Missouri 63042-2320

Distribution

Distribution pattern
Worldwide Distribution - USA including AR, AZ, CA, FL, KS, LA, MD, MN, MT, NC, NV, NY, OR, PA, RI, SC, TX, VA and WA. Internationally to Argentina, Armenia, Austria, Bangladesh, Bosnia and Herzegowina, Brazil, Bulgaria, Burkina FASO, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, COTE DIVOIRE, CROATIA, CZECH REPUBLIC, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FRANCE, FRENCH POLYNESIA, GERMANY, GREECE, GUATEMALA, HONDURAS, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LITHUANIA, MALTA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PANAMA, PERU, POLAND, PORTUGAL, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, UNITED STATES, URUGUAY, VENEZUELA, VIETNAM, PALESTINE, AND SERBIA.

Timeline

Recall initiated
2012-10-16
Posted by FDA
2012-11-16
Terminated
2013-10-28
Status

Source: openFDA Device Recall endpoint. Recall record ID #114016. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.