Recalls / —
—#114076
Product
UniCel DxC 880i, 860i, 680i, and 660i Integrated Systems, Part Numbers: A64871, A64903, A64935, A59102. The UniCel DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The UniCel DxC series of instruments is a DxI Immunoassay System integrated with a chemistry analyzer.
- FDA product code
- CEM — Electrode, Ion Specific, Potassium
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K042291
- Affected lot / code info
- Serial Numbers: Included in the corresponding Dxl system.
Why it was recalled
The recall was initiated because Beckman Coulter has confirmed that the UniCel Dxl and DxC Access Immunoassay Systems had field modifications to the aspirate and dispense probe assemblies.
Root cause (FDA determination)
Process design
Action the firm took
Beckman Coulter sent a field modification was released in August 2011 which included: - Installing a new aspirate and dispense tubing routing bracket. - Replacing existing aspirate and dispense probe tubing with modified tubing to help eliminatemisrouting of aspirate and dispense probe tubing. During installation the FSE informed the customer: -That aspirate and dispense tubing are different lengths. - That probes are not interchangeable with other instruments once tubing has been modified. - To route probe tubing by referring to the labels on the tubing. An updated service document, Routine Maintenance Changes, A82294B was provided to every customer. For questions regarding this recall call 714-961-4483.
Recalling firm
- Firm
- Beckman Coulter Inc.
- Address
- 250 S Kraemer Blvd, Brea, California 92821-6232
Distribution
- Distribution pattern
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, AE, AU, BG, BW, CH, CN, CZ, DE, ES, FR, GB, HK, HR, HU, IN, IT, JO, JP, LB, MA, MX, NL, PL, QA, RO, RU, SK, TN, TR, ZA, AO, AR, AT, BE, CO, DZ, FI, IE, IL, KR, KW, LI, NA, NO, NZ, PF, PH, PR, SE, SG, SZ, TH, TW, VE, VN, and YT.
Timeline
- Recall initiated
- 2011-08-01
- Posted by FDA
- 2012-11-19
- Terminated
- 2012-11-19
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #114076. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.