FDA Device Recalls

Recalls /

#114124

Product

00789502601 VerSys Femoral Head Provisional 26mm -3.5 00789502602 VerSys Femoral Head Provisional 26mm +0 00789502603 VerSys Femoral Head Provisional 26mm +3.5 00789502604 VerSys Femoral Head Provisional 26mm +7 00789502605 VerSys Femoral Head Provisional 26mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.

FDA product code
HWTTemplate
Device class
Class 1
Medical specialty
Orthopedic
Affected lot / code info
All lots distributed from lot 69998600 (2/20/1996) to present

Why it was recalled

Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat

Root cause (FDA determination)

Device Design

Action the firm took

The firm, Zimmer, sent notification by email to consignees beginning October 10, 2012 with a follow-up letter titled "URGENT MEDICAL DEVICE CORRECTION" dated October 16, 2012 to its consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to complete a spreadsheet of current users and contact information and immediately provide this information via email to: CorporateQuality.Postmarket@zimmer.com, and review and become familiar with the content and recommendations of the attached letter sent to the customers. For questions call 1-877-946-2761.

Recalling firm

Firm
Zimmer, Inc.
Address
345 E Main St, Warsaw, Indiana 46580-2746

Distribution

Distribution pattern
Worldwide distribution: USA (nationwide) and countries of: Asia, Barbados, Canada, Europe, Mexico, Puerto Rico, San Juan, and South America.

Timeline

Recall initiated
2010-10-10
Posted by FDA
2012-11-19
Terminated
2014-04-30
Status

Source: openFDA Device Recall endpoint. Recall record ID #114124. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.