Recalls / —
—#114387
Product
Access Immunoassay System, Access 2 Immunoassay System, Synchron LXi 725 Clinical System, UniCel DxC 600i Synchron Access Clinical System Part Numbers: Access 81600; Access 2 81600N, 386220, A69186, A25656, A12900; Access 2 (LXi) A15642; Access 2 Section, DxC 600i A25640. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The Synchron LXi and the UniCel DxC 600i are Access systems integrated with a chemistry analyzer.
- FDA product code
- JJE — Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Device class
- Class 1
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K023049, K060256, K922823
- Affected lot / code info
- Part Numbers: Access 81600; Access 2 81600N, 386220, A69186, A25656, A12900; Access 2 (LXi) A15642; Access 2 Section, DxC 600i A25640
Why it was recalled
It has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at an incorrect speed, resulting in inadequate washing or mixing of the contents of the reaction vessel.
Root cause (FDA determination)
Device Design
Action the firm took
Beckman Coulter sent an "URGENT PRODUCT CORRECTION" letter dated November 1, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were asked to complete and return an enclosed Response Form. Contact Technical Support for questions regarding this notice.
Recalling firm
- Firm
- Beckman Coulter Inc.
- Address
- 250 S Kraemer Blvd, Brea, California 92821-6232
Distribution
- Distribution pattern
- Worldwide Distribution-USA (nationwide) and the country of Canada.
Timeline
- Recall initiated
- 2012-11-01
- Posted by FDA
- 2012-12-22
- Terminated
- 2013-12-23
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #114387. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.