FDA Device Recalls

Recalls /

#114526

Product

Zimmer Spine Ardis Inserter, Rx, non-sterile

FDA product code
MAXIntervertebral Fusion Device With Bone Graft, Lumbar
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K073202
Affected lot / code info
Ardis Inserter, Part number: 3256-01, Lot: 52ZW, 50CL, 53BB, 55AB, 58GS, 58GT, 60JY, 60JZ, 62KP, 63HJ, 64EY, 68AX, 71JE, 75LE   Part, 07.00780.701, lot 76CN; part 07.00780.702, lot 76CP;part 07.00900.701, lot 76TR; 07.00900.702, 76TS; 650-0297-01, 70YT; 07.00780.701, 73EC; 07.00780.702, 73EC; 07.00900.701, 76TQ; 650-0239-01, 66PR; 650-0293-01, 70SG ; 650-0293-01, 70SG-R; 650-0179-01, 60CK; 650-0176-01, 60CL; 650-0365-01, 72LP; 07.00780.701, 76CQ; 07.00780.702, 76CR; 07.00780.701, 74EN; 07.00780.702, 74EP; 650-0351-01, 72JH; 650-0351-02, 72JJ; 650-0263-01, 68HL; 650-0208-01, 62FH; 07.00900.701, 76TN; 07.00780.701, 76QG; 07.00780.701, 76LC; 07.00780.702, 76LD;

Why it was recalled

Zimmer Spine has received reports of the PEEK Ardis Interbody Spacer breaking into fragments when too much lateral force is applied to the Ardis Inserter. If the PEEK Ardis Interbody Spacer breaks during implantation, surgical delays of up to 60 minutes could occur. Sharp edges of the broken implant may contact and damage adjacent structures, requiring surgical intervention. Observed health risks

Root cause (FDA determination)

Device Design

Action the firm took

Zimmer Spine, sent a "MEDICAL DEVICE CORRECTION" memo dated November 12, 2012 to all consignees / customers describing product, problem and actions to be taken including revised surgical technique. Beginning the week of December 17, 2012, Zimmer began the immediate retrieval of the affected units as a Class I recall. December 20, 2012 nation wide press outlining the recall was issued. Customers may report adverse events or quality problems experienced with the use of these products to Zimmer Spine by calling 866-774-6368 between the hours of 8AM and 5PM CST or by email at qualitycompliance@zimmer.com.

Recalling firm

Firm
Zimmer, Inc.
Address
345 E Main St, Warsaw, Indiana 46580-2746

Distribution

Distribution pattern
Worldwide distribution: USA (nationwide) including states of: AR, AZ, CA, CO, CT, DC, FL, GA, ID, LA, MD, MI, MN, MO, NC, NV, NY, OH, OK, PA, TN, TX, WI, and WV; and countries of: Australia, El Salvador, France and Singapore.

Timeline

Recall initiated
2012-11-12
Posted by FDA
2012-12-22
Terminated
2014-02-25
Status

Source: openFDA Device Recall endpoint. Recall record ID #114526. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.