Product

Siemens Chek-Stix Combo Pak (Distributed OUS only) Catalog Number: 1364M Chek-Stix Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes

FDA product code
JJWUrinalysis Controls (Assayed And Unassayed)
Device class
Class 1
Medical specialty
Clinical Chemistry
510(k) numbers
K943780
Affected lot / code info
Carton/Bottle Lot Number: CP0099091B P0099091B C0100101G P0100101G C0103042E P0103042E

Why it was recalled

When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F).

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Siemens issued An Urgent Field Safety Notice via e-mail to all affected SiemensHealthcare Diagnostics Regional offices on November 21, 2012 both in the United States and foreign accounts. This notice informed customers of the issue with the CLINITEK Atlas and Chek-Stix Positive Urinalysis Control Solutions When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F). You can use the reconstituted positive control solution for up to eight hours if the solution is held at 25C (77F) or lower. If the operating temperature is greater than 25C and lower than 30C (77F and 86F), the solution can be used for up to two hours. For questions regarding this recall call 781-269-3158.

Recalling firm

Firm
Siemens Healthcare Diagnostics Inc
Address
2 Edgewater Dr, Norwood, Massachusetts 02062-4637

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,

Timeline

Recall initiated
2012-11-21
Posted by FDA
2013-01-04
Terminated
2014-11-06
Status