Recalls / —
—#114833
Product
Dade(R) Actin(R) FSL Activated PTT Reagent. For use in the determination of the activated partial thromboplastin time (APTT).
- FDA product code
- GGW — Test, Time, Partial Thromboplastin
- Device class
- Class 2
- Medical specialty
- Hematology
- 510(k) numbers
- K863594
- Affected lot / code info
- Catalog numbers B4219-1 and B4219-2: lot numbers 547311, exp 3/22/2013; 547312, exp 3/22/2013 and 547316A, exp 4/28/2013.
Why it was recalled
Siemens has observed that the listed lot numbers show an increase in Heparin sensitivity over the shelf life which is demonstrated by testing with Dade(R) Citrol(R) Heparin Control, low and high.
Root cause (FDA determination)
Unknown/Undetermined by firm
Action the firm took
Siemens sent an "URGENT FIELD SAFETY NOTICE" dated February 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The safety notice instructed consignees to discontinue use of the affected lots and discard any remaining inventory. Contact the firm at 1-800-242-3233, option 1 for questions about this notice.
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Address
- 500 Gbc Dr PO BOX 6101, Ms 514, Newark, Delaware 19702-2466
Distribution
- Distribution pattern
- Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV and Puerto Rico and the country of Canada.
Timeline
- Recall initiated
- 2013-03-11
- Posted by FDA
- 2013-04-09
- Terminated
- 2013-12-24
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #114833. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.