Recalls / —
—#114845
Product
Cordis, a Johnson and Johnson Company, ExoSeal(TM) 6F Vascular Closure Device, cat. No EX500, EX600, and EX700. Used for femoral artery puncture site closure, reducing times to hemostasis and ambulation in patients who have undergone diagnostic or interventional procedures.
- FDA product code
- MGB — Device, Hemostasis, Vascular
- Device class
- Class 3
- Medical specialty
- Unknown
- PMA numbers
- P100013
- Affected lot / code info
- Model EX500: Lot #'s: 15410906, 15411659, 15424968, 15429098, 15429879, 15437509, 15437958, 15444208, 15449422, 15453540, 15456567, 15456568, 15462779, 15462780, 15469984, 15479199, 15481341, 15485491, 15489970, 15501955, 15505186, 15578709, 15589368, 15586530, 15602199, 15603369, 15608590, 15610801, 15617548, 15630141, 15630171, 15639388, 15642692, 15648662, 15650923, 15656441, 15664083, 15670351, 15670768. Model EX600: Lot #'s: 15408606, 15409332, 15410622, 15410623, 15426890, 15430798, 15433095, 15433098, 15441594, 15443327, 15446063, 15448005, 15451542, 15452987, 15455770, 15455771, 15459366, 15459367, 15461179, 15461781, 15463296, 15463297, 15464298, 15464299, 15466362, 15472044, 15475128, 15475795, 15479644, 15484135, 15487458, 15491737, 15496311, 15497038, 15500078, 15500709, 15503015, 15504223, 15504224, 15504748, 15506926, 15508012, 15508353, 15508856, 15510541, 15512713, 15518232, 15540529, 15545067, 15548555, 15549324, 15550196, 15550197, 15557284, 15561545, 15561546, 15566744, 15572185, 15573480, 15579468, 15585121, 15587184, 15596793, 15601234, 15601753, 15604477, 15605089, 15609335, 15613282, 15617549, 15625033, 15627316, 15635233, 15637262, 15643624, 15644116, 15644789, 15644790, 15649759, 15663330. Model EX700 Lot #'s: 15466361, 15471271, 15471272, 15476317, 15476743, 15482155, 15483371, 15486224, 15490397, 15535011, 15540530.
Why it was recalled
Affected lots of Cordis ExoSeal (TM) VCD have been sterilized using a process that did not achieve the sterility assurance level that is required by the standards of this class of product. Model Numbers EX500, EX600, and EX700.
Root cause (FDA determination)
Material/Component Contamination
Action the firm took
Cordis Corporation sent an "Urgent Medical Device Recall (Removal ) letter dated October 29, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to immediately identify and set aside all product listed in a manner that ensures the affected product will not be used. A product recall response form was attached for customers to review, complete, sign and return the enclosed acknowledgement form directly to Cordis. For further questions please call (786) 313-2000.
Recalling firm
- Firm
- Cordis Corporation
- Address
- 14201 Nw 60th Ave, Miami Lakes, Florida 33014-2802
Distribution
- Distribution pattern
- US (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and Hawaii.
Timeline
- Recall initiated
- 2012-10-29
- Posted by FDA
- 2013-01-25
- Terminated
- 2013-11-06
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #114845. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.