Product

Cordis, a Johnson and Johnson Company, ExoSeal(TM) 6F Vascular Closure Device, cat. No EX500, EX600, and EX700. Used for femoral artery puncture site closure, reducing times to hemostasis and ambulation in patients who have undergone diagnostic or interventional procedures.

FDA product code
MGBDevice, Hemostasis, Vascular
Device class
Class 3
Medical specialty
Unknown
PMA numbers
P100013
Affected lot / code info
Model EX500: Lot #'s: 15410906, 15411659, 15424968, 15429098, 15429879, 15437509, 15437958, 15444208, 15449422, 15453540, 15456567, 15456568, 15462779, 15462780, 15469984, 15479199, 15481341, 15485491, 15489970, 15501955, 15505186, 15578709, 15589368, 15586530, 15602199, 15603369, 15608590, 15610801, 15617548, 15630141, 15630171, 15639388, 15642692, 15648662, 15650923, 15656441, 15664083, 15670351, 15670768.  Model EX600: Lot #'s: 15408606, 15409332, 15410622, 15410623, 15426890, 15430798, 15433095, 15433098, 15441594, 15443327, 15446063, 15448005, 15451542, 15452987, 15455770, 15455771, 15459366, 15459367, 15461179, 15461781, 15463296, 15463297, 15464298, 15464299, 15466362, 15472044, 15475128, 15475795, 15479644, 15484135, 15487458, 15491737, 15496311, 15497038, 15500078, 15500709, 15503015, 15504223, 15504224, 15504748, 15506926, 15508012, 15508353, 15508856, 15510541, 15512713, 15518232, 15540529, 15545067, 15548555, 15549324, 15550196, 15550197, 15557284, 15561545, 15561546, 15566744, 15572185, 15573480, 15579468, 15585121, 15587184, 15596793, 15601234, 15601753, 15604477, 15605089, 15609335, 15613282, 15617549, 15625033, 15627316, 15635233, 15637262, 15643624, 15644116, 15644789, 15644790, 15649759, 15663330.  Model EX700 Lot #'s: 15466361, 15471271, 15471272, 15476317, 15476743, 15482155, 15483371, 15486224, 15490397, 15535011, 15540530.

Why it was recalled

Affected lots of Cordis ExoSeal (TM) VCD have been sterilized using a process that did not achieve the sterility assurance level that is required by the standards of this class of product. Model Numbers EX500, EX600, and EX700.

Root cause (FDA determination)

Material/Component Contamination

Action the firm took

Cordis Corporation sent an "Urgent Medical Device Recall (Removal ) letter dated October 29, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to immediately identify and set aside all product listed in a manner that ensures the affected product will not be used. A product recall response form was attached for customers to review, complete, sign and return the enclosed acknowledgement form directly to Cordis. For further questions please call (786) 313-2000.

Recalling firm

Firm
Cordis Corporation
Address
14201 Nw 60th Ave, Miami Lakes, Florida 33014-2802

Distribution

Distribution pattern
US (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and Hawaii.

Timeline

Recall initiated
2012-10-29
Posted by FDA
2013-01-25
Terminated
2013-11-06
Status