FDA Device Recalls

Recalls /

#115133

Product

Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Adhesive Area (Cat. No. 4FA40SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 4FM40SSA) Product Usage: Disposal system for counting of used surgical blades and needles.

FDA product code
FSMTray, Surgical, Instrument
Device class
Class 1
Medical specialty
General, Plastic Surgery
Affected lot / code info
Catalog Number: FA40SSA Lot Numbers: lot number 763998  Catalog Number: 4FM40SSA Lot Numbers: all lot numbers less than or equal to 933099

Why it was recalled

Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.

Root cause (FDA determination)

Package design/selection

Action the firm took

Cardinal Health sent an Urgent Product Recall letter dated December 18, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immedaitely stop using affected product and discard any product that may be on hand, return the enclosed acknowledgement form, notify any customers to whom you may have distributed product affected by this recall, and contact the appropriate Customer Service group to arrange for credit of any affected product discarded. For questions contact Cardinal Health Professional Services at 800-292-9332.

Recalling firm

Firm
Cardinal Health
Address
1430 Waukegan Rd, Mc Gaw Park, Illinois 60085-6726

Distribution

Distribution pattern
Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.

Timeline

Recall initiated
2012-12-18
Posted by FDA
2013-03-28
Terminated
2016-08-01
Status

Source: openFDA Device Recall endpoint. Recall record ID #115133. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable N · FDA Device Recalls