Recalls / —
—#115545
Product
Merit Medical Systems Inc., Resolve Biliary Locking Drainage Catheter, Models RBC and RBDC, used for drainage of bile within the biliary system.
- FDA product code
- FGE — Stents, Drains And Dilators For The Biliary Ducts
- Device class
- Class 2
- Medical specialty
- Gastroenterology, Urology
- 510(k) numbers
- K121832
- Affected lot / code info
- Lot numbers H399931S1; H411376; H399938S1; H411373; H399940S1; H411374; H399942S1; H411375; H399948S1; H411380; H399950S1; H411377; H39952S1; H411378; H39953S1; H411379.
Why it was recalled
A consumer complaint that a catheter was observed to be in several pieces under patient fluoroscopic visualization.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
Merit initiated the recall on January 16, 2013. A voicemail script notification to users and Q&A document was provided to those sales representative that had received and or distributed product. On January 16, 2013, Merit sales representatives began contacting each physician and/or user facility that had been provided product. Written communication distribution began on 22 January 2013. Affected sales representatives, physicians, and hospitals were provided written identification information, instruction to not distribute or continue placing these devices, and on how to return any unused product if applicable. All product on hand at Merit was inventoried and remains under quarantined status. Further questions please call (801) 208-4301,
Recalling firm
- Firm
- Merit Medical Systems, Inc.
- Address
- 1600 West Merit Parkway, South Jordan, Utah 84095
Distribution
- Distribution pattern
- Worldwide Distribution - USA including the states of NJ, FL, GA, MO, AZ, UT and the country of Denmark
Timeline
- Recall initiated
- 2013-01-16
- Posted by FDA
- 2013-02-20
- Terminated
- 2015-09-16
- Status
- —