FDA Device Recalls

Recalls /

#115642

Product

ACUSON S2000 Automated Breast Volume Scanner (ABVS); Model number 10042541 Distributed by Siemens Medical Solutions USA Inc. business Unit: Ultrasound. The S2000 Automated Breast Volume Scanner (ABVS) Ultrasound System for imaging of the breast allows volumetric acquisition of the breast image.

FDA product code
IYOSystem, Imaging, Pulsed Echo, Ultrasonic
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K081148
Affected lot / code info
Model number 10042541 - AVBS module.  Note: the S2000 system part number is 10041461, however only those systems that include the ABVS module, part number 10042541 are affected. Manufactured prior to 10/30/2012.

Why it was recalled

On some ACUSON S2000 ABVS systems, the inside of the transducer pod may contain sharp edges.

Root cause (FDA determination)

Process design

Action the firm took

Siemens sent a Customer Safety Advisory notification letter to Siemens Service Personnel on January 18, 2013 and which would be provided to all affected sites during their visit to correct the problem. The notification identified the product, the problem and the action to be taken. The local Customer Service Engineer will perform the test at their site. If the system passes the test, no further action is required. If the ABVS pod fails the inspection, the Service Engineer will correct the sharp edge. For questions regarding this recall call 650-694-5398.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc.
Address
685 E Middlefield Rd, Mountain View, California 94043-4045

Distribution

Distribution pattern
Worldwide Distribution - USA including WV and OH; Internationally to Australia, Austria, Bahrain, Belgium, Brazil, China (P.R.) Estonia, France, Germany, Greece, India, Indonesia, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Macedonia, Malaysia, Mexico, Netherlands, Peru, Philippines, Poland, Portugal, Qatar, Rep of Korea, Romania, Sweden, Taiwan, Thailand, Turkey, Unite Arab Emirates, United Kingdom and Venezuela.

Timeline

Recall initiated
2013-01-18
Posted by FDA
2013-03-09
Terminated
2013-07-16
Status

Source: openFDA Device Recall endpoint. Recall record ID #115642. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.