FDA Device Recalls

Recalls /

#115662

Product

ELECTROLASE HYFRECATOR(R) ELECTRODES - SHARP, REF/Catalog Number 7-100-8BX, Disposable Hyfrecator(R) Tips, Sterile, Rx Only --- CONMED CORPORATION USA. Intended to be used as active RF electrodes for desiccation, fulguration and coagulation.

FDA product code
GEIElectrosurgical, Cutting & Coagulation & Accessories
Device class
Class 2
Medical specialty
General, Plastic Surgery
510(k) numbers
K800617
Affected lot / code info
Beginning Lot Code 120628X, Ending Lot Code 120917X. [Lot codes on boxes and packaging contain a lot code in the following format: 120917X where 12 represents the year (2012), 09 represents the month (September), 17 represents the day of the month, and X represents the manufacturing shift code (1, 2, 3, 4,5,6 or 9)].

Why it was recalled

Some devices had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products.

Root cause (FDA determination)

Equipment maintenance

Action the firm took

ConMed sent Urgent Device Recall Letters (dated 1/28/13) and Business Reply Forms to all domestic customers via USPS Priority Mail. International notices were sent via FedEx International Priority to the foreign consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory and complete Attachment II response form and return it with unused affected product to: ConMed Corporation 525 French Road, Utica, NY 13502. Customers were asked to return Attachment II even if they don't have any affected product. For questions regarding this recall call 315-624-3533.

Recalling firm

Firm
ConMed Corporation
Address
525 French Road, Utica, New York 13502

Distribution

Distribution pattern
Worldwide Distribution-USA (nationwide) and the countries of China, Dubai and Indonesia.

Timeline

Recall initiated
2013-01-28
Posted by FDA
2013-02-15
Terminated
2017-04-11
Status

Source: openFDA Device Recall endpoint. Recall record ID #115662. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.