FDA Device Recalls

Recalls /

#116027

Product

VITEK 2 Compact computer system when equipped with the following Hewlett-Packard PC models: (1) dc7100 8-port, Cat. #W0441; (2) dc7100 8-port (refurbished), Cat. #W0441R; (3) dc7700 8-port, Cat. #W0447; (4) dc7700 8-port (refurbished), Cat. #W0447R; (5) dc7800, Cat. #W0449; (6) dc7800 (refurbished), Cat. #W0449R (7) rp5700, Cat. #W0452; (8) rp5700 (refurbished), Cat. #W0452R; (9) rp5800 XPE, Cat. 413642; (10) rp5800WES7, Cat. #413862; (11) d530 6-port, Cat. #W0438; (12) d530 6-port (refurbished), Cat. #W0438R; and (13) d530 4-port, Cat. #W0437. VITEK 2 is an automated system designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.

FDA product code
LTWSusceptibility Test Cards, Antimicrobial
Device class
Class 2
Medical specialty
Microbiology
510(k) numbers
K050002
Affected lot / code info
Manufactured from September, 2004, to present.

Why it was recalled

There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.

Root cause (FDA determination)

Software design

Action the firm took

bioMerieux sent an "URGENT PRODUCT CORRECTION NOTICE" dated February 20, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Emails were issued by the recalling firm on 2/18/2013 and 2/19/2013 to their subsidiaries and to distributors. Letters began issuing to customers via e-mail on 2/20/2013. Contact bioMerieux Clinical Customer Service at (800) 682-2666 for questions regarding this notice.

Recalling firm

Firm
Biomerieux Inc
Address
595 Anglum Rd, Hazelwood, Missouri 63042-2320

Distribution

Distribution pattern
Worldwide Distribution-USA (nationwide) and the countries of Algeria, Angola, Argentina, Australia, Austria, Azerbaidjian, Bahrain, Belarus, Belgium, Benin, Bolivia, Bosnia, Brazil, Bulgaria, Burkina Faso, Camaroun, Canada, Chad, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Democratic Republic of Congo, Demark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Guinea, Herzegovina, Honduras, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mali, Mexico, Moldavia, Mongolia, Morocco, Myanmar, Netherlands, Netherlands, Nicaragua, Niger Niamey, Nigeria, Norway, Oman, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Yemen.

Timeline

Recall initiated
2013-02-18
Posted by FDA
2013-03-25
Terminated
2014-08-14
Status