Recalls / —
—#116367
Product
Philips IntellVue Info Center iX 866023; PIIC iX Upgrade 866024; IntellVue Server iX Upgrade 866026; PIIC Classic Upgrade 866117 Patient Physiological Monitor.
- FDA product code
- MHX — Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K102495
- Affected lot / code info
- All iX versions with A.01 software
Why it was recalled
If a customer creates customized trend scales in the trend review tile and the iX or primary server reboots for any reason, the iX(s) will enter a reboot loop. If one iX reboots, the issue will impact only those patients monitored on that device. If the primary server reboots, all patients across all iX devices will be impacted.
Root cause (FDA determination)
Software design
Action the firm took
On 3/5/13, Philips sent each known affected customer an Urgent Medical Device Correction notification/Field Safety Notice to inform them that Philips has received a customer report that the PIC iX (release A.01.XX) central stations that are connected to a database server (DBS) may not be able to complete a restart/reboot resulting in the loss of central monitoring. In order to prevent this from occurring, customers were advised to not modify/customize the default trend scales for reviewing stored patient data in the retrospective review applications. Recall letter also states that a Philips Healthcare representative will contact customers with affected devices to arrange for the installation of updated PIIX iX software resolving this issue on affected units. Philips will conduct these updates for all affected devices at no charge.
Recalling firm
- Firm
- Philips Healthcare Inc.
- Address
- 3000 Minuteman Road, Andover, Massachusetts 01810
Distribution
- Distribution pattern
- Worldwide Distribution: US (nationwide) and country of: Canada.
Timeline
- Recall initiated
- 2013-03-05
- Posted by FDA
- 2014-06-11
- Terminated
- 2017-05-17
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #116367. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.