Recalls / —
—#116407
Product
Sarns " TCM II Temperature Control Module The Sarns" TCM II Temperature Control Module is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and is indicated for controlling and monitoring patient temperature
- FDA product code
- DWC — Controller, Temperature, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K883603
- Affected lot / code info
- Catalog Number 4416 and Serial Numbers: 4656, 4657, 4658,4659, 4661, 4662, and 4663.
Why it was recalled
Internal testing found certain Sarns" TCM II systems were released with incorrectly assembled heatsink assemblies which could cause the units to fail.
Root cause (FDA determination)
Employee error
Action the firm took
Terumo sent beginning May 30, 2013, an URGENT MEDICAL DEVICE RECALL NOTIFICATION were sent to all customers with instructions to notify all customers of the recall requesting all affected units to be quarantined until the heatsink units are replaced Confirm receipt by filling out the response form containing communication by faxing or emailing the attached Affiliate Response Form to the fax number or email address indicated on the form. Questions should be directed to Terumo CVS customer service in the U.S.: Telephone: +1-734-663-4145; Fax: +1-734-741-6149
Recalling firm
- Firm
- Terumo Cardiovascular Systems Corporation
- Address
- 6200 Jackson Road, Ann Arbor, Michigan 48103-9586
Distribution
- Distribution pattern
- Worldwide Distribution including the countries of UAE (United Arab Emirates) HONG KONG and THAILAND. Not distributed in USA
Timeline
- Recall initiated
- 2013-05-30
- Posted by FDA
- 2013-06-21
- Terminated
- 2013-09-17
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #116407. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.