Recalls / —
—#116490
Product
CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").
- FDA product code
- GJS — Test, Time, Prothrombin
- Device class
- Class 2
- Medical specialty
- Hematology
- 510(k) numbers
- K093460
- Affected lot / code info
- catalog number 05530199160
Why it was recalled
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.
Root cause (FDA determination)
Device Design
Action the firm took
Consignees of both devices were notified by Urgent Medical Device Correction letters sent via UPS Ground (receipt required) beginning 3/06/2013. The notice contained recommendation not to use EAN-13 barcode system and ensure barcodes used meet quality standards; use operator and patient lists (downloaded from a data manager) in order to check any decoded identification against the respective list. The notice directed forwarding of the notice if the product was further distributed. Customers were asked to complete the attached fax form. Contact the Customer Care Service Center, 24 hours a day, seven days a week at 1-800-440-3638 for Accu check questions and 1-800-428-4674 for CoaguChek questions about the information contained in this UMDC.
Recalling firm
- Firm
- Roche Diagnostics Operations, Inc.
- Address
- 9115 Hague Rd, Indianapolis, Indiana 46256-1025
Distribution
- Distribution pattern
- Nationwide Distribution.
Timeline
- Recall initiated
- 2013-03-06
- Posted by FDA
- 2013-04-15
- Terminated
- 2018-01-10
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #116490. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.