Recalls / —
—#116754
Product
Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Select. Angiographic x-ray system.
- FDA product code
- JAA — System, X-Ray, Fluoroscopic, Image-Intensified
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K992660
- Affected lot / code info
- Model Number 5902767 with serial numbers 2217, 2038, 1228, 1007, 1219, 2156, 2264, 2190, 2132, 2213, 1484, 2166, 2259, 2068, 1610, 1601, 1413, 2236, 2153, 1444, 1477, 1632, 2037, 2180, 1663, and 1612; Model Number 5902783 with serial numbers 1003, 1401, 1402, and 1405; Model Number 7152478 with serial numbers 2580, 2524, 2590, 2536, 2563, and 2534; Model Number 10093962 with serial numbers 5837, 5120, 5523, 5910, 5507, 6141, 4112, 6274, 4506, 5900, 4121, 6245, 6242, 4509, and 6113.
Why it was recalled
Siemens has become aware of a potential malfunction and possible hazard to patients and staff when using the Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Select. This malfunction is possible in cases where the table vertical lift movement is obstructed by a solid object which could potentially result in collision. This can re
Root cause (FDA determination)
Process design
Action the firm took
Siemens sent a Customer Safety Advisory letter dated January 15th, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The SIEMENS service representative will investigate if parts need to be exchanged in your product, an if so, which parts need to be exchanged to restore the system safety and function. Please kindly confirm receipt of the Field Safety Notice by sending back the attached form. For further questions please call (610) 448-3237.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 51 Valley Stream Pkwy, Malvern, Pennsylvania 19355
Distribution
- Distribution pattern
- Nationwide Distribution including the states of CA, FL, GA, IA, IN, KS, LA, ME, MI, MO, NC, NH, NJ, NY, NV, OH, PA, TN, TX, VA and WA.
Timeline
- Recall initiated
- 2013-01-15
- Posted by FDA
- 2013-04-11
- Terminated
- 2013-04-23
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #116754. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.