Recalls / —
—#116762
Product
Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
- FDA product code
- GCX — Apparatus, Suction, Operating-Room, Wall Vacuum Powered
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K954726
- Affected lot / code info
- Manufacturer Part Number: 250-070-403
Why it was recalled
The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123C , instead of the correct 132-133C.
Root cause (FDA determination)
Error in labeling
Action the firm took
Stryker sent "Urgent Device Correction" letters dated March 27, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Recalling firm
- Firm
- Stryker Endoscopy
- Address
- 5900 Optical Ct, San Jose, California 95138-1400
Distribution
- Distribution pattern
- Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.
Timeline
- Recall initiated
- 2013-03-27
- Posted by FDA
- 2013-05-03
- Terminated
- 2013-09-24
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #116762. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.