Product

Stryker Irrigator Reusable Tips: 2mm X 34cm Regular Tip with No Hole, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.

FDA product code
GCXApparatus, Suction, Operating-Room, Wall Vacuum Powered
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K954726
Affected lot / code info
Manufacturer Part Number: 250-070-473

Why it was recalled

The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123C , instead of the correct 132-133C.

Root cause (FDA determination)

Error in labeling

Action the firm took

Stryker sent "Urgent Device Correction" letters dated March 27, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.

Recalling firm

Firm
Stryker Endoscopy
Address
5900 Optical Ct, San Jose, California 95138-1400

Distribution

Distribution pattern
Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.

Timeline

Recall initiated
2013-03-27
Posted by FDA
2013-05-03
Terminated
2013-09-24
Status

Source: openFDA Device Recall endpoint. Recall record ID #116782. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.