Product
MAQUET CARDIOHELP Support System: CARDIOHELP-I Software (up to version 3.3.0.0) Product Usage: The Cardiohelp System is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support during procedures requiring cardiopulmonary bypass (for periods up to six hours). It is also intended to provide circulatory and/or pulmonary support during procedures not requiring cardiopulmonary bypass (for periods up to six hours).
- FDA product code
- DTQ — Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K102726
- Affected lot / code info
- Catalog Number: 70104.8012 Cardiohelp-i. Multiple serial numbers.
Why it was recalled
It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touch screen) of the CARDIOHELP device. The human machine interface (HMI) is the central display of the CARDIOHELP device, where measured values, settings and alarms are displayed and adjusted/confirmed, with the exception of bl
Root cause (FDA determination)
Software change control
Action the firm took
MAQUET sent an Urgent Medical Device Field Correction letters dated March 7, 2013 via Federal Express to all affected customers. The letter identifies the affected product, problem and actions to be taken. Customers are asked to complete and submit the included Customer Fax-back Form. Maquet Service will exchange all affected products with the updated version. For questions MAQUET Customer Service at 888-627-8383 (press option 2, followed by option 2).
Recalling firm
- Firm
- Maquet Cardiovascular, LLC
- Address
- 45 Barbour Pond Dr, Wayne, New Jersey 07470-2094
Distribution
- Distribution pattern
- Worldwide Distribution - USA Nationwide and the following countries: Australia, Austria, Belarus, Belgium, Canada, Columbia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Iceland, Iran, Ireland, Italy, Jordan, Kazakhstan, Korea, Kuwait, Luxembourg, The Netherlands, Norway, Oman, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates and the United Kingdom.
Timeline
- Recall initiated
- 2013-03-07
- Posted by FDA
- 2013-04-18
- Terminated
- 2013-08-22
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #116845. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.