FDA Device Recalls

Recalls /

#117262

Product

Optima MR430s MRI Scanner; 5000-0002 is indicated to use as a magnetic resonance imaging device of the leg (excluding the thigh), knee, ankle, foot, elbow, forearm, wrist, and hand.

FDA product code
LNHSystem, Nuclear Magnetic Resonance Imaging
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K103238
Affected lot / code info
08022011-001; 09092011-001;12102011-001;05152012-004;SM11062012-001;02022012-001;08142012-001;04212012-001;05252012-002;12132011-001;SM08292012-007;11052011-002;01192012-001;01132012-001;06282012-010;07252012-001;01022012-001;11202011-001;08142011-001;06302011-001;09172011-001;10082011-001;10172011-002;03142012-007;06142012-001;06212012-001;07172012-001;07232012-001;SM09132012-003;02042012-001;01132012-001;07172011-001;09062011-001;09242011-001;SM10122012-003;M12102012-003;05102012-004;08022012-003;07302012-008;SM01072013-005;05032012-004;08032012-001;10172011-001;SM08152012-001;12262011-001;10092011-001;04172012-001;11052011-001;07112011-001;10242011-007;07222011-001;08102011-001;03092012-001;05172011-001;06222011-001;03202012-004;05182012-001;07112012-002;06092011-001;01232012-001;02162012-001;11262012-007;SM12172012-001;01212012-001;SM10232012-001;11212011-001;12222011-001;SM11132012-002;06012011-001;07132011-001;12042011-001;03062012-007;03282012-001;10272011-009;12142011-001;10302012-006;02112012-001;03062012-008;01272012-001;05162012-003;03262012-002;11292011-001;06112012-004;10172012-001;12102012-001

Why it was recalled

Coolant supply lines may leak liquid onto the floor resulting in a slippery surface and fall hazard.

Root cause (FDA determination)

Process control

Action the firm took

GE Healthcare sent an Urgent Medical Device Correction letter dated April 23, 2013, to all affected customers. The letter informed customers of a potential safety issue due to leakage from gradient coolant lines. Consignees were instructed to monitor the floor for leakage and continue to use if leakage is not observed. If leakage is observed consignees are instructed to stop using the system and notify their local service representative. GE Healthcare will correct all affected scanners by providing an update at no cost. Customers with questions were instructed to contact their local service representataive. For questions regarding this recall call 262-513-4122.

Recalling firm

Firm
GE Healthcare, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
Worldwide Distribution - USA including AL, CA, CO, FL, ID, IL, MA, NH, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI; Worldwide: Argentina, Australia, Austria, Brazil, Canada, Chile, Denmark, Egypt, Finland, Germany, India, Italy, Japan, Korea, Malaysia, Netherlands, Norway, Peru, Poland, Russia, Singapore, South Africa, Spain, Sweeden, Switzerland, UK, Ukraine, and UAE.

Timeline

Recall initiated
2013-03-28
Posted by FDA
2013-05-01
Terminated
2013-10-25
Status

Source: openFDA Device Recall endpoint. Recall record ID #117262. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.