Recalls / —
—#117800
Product
Dimension(R) TACR Flex(R) reagent cartridge (DF107) The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension(R) clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of Tacrolimus therapy in kidney and liver transplant patients
- FDA product code
- MLM — Enzyme Immunoassay, Tracrolimus
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- Affected lot / code info
- Siemens Material Number 10444938, lot numbers GB3099, GA3120, DB3141, GB3162, GB3176, FA3197, FA3267 and FB3267.
Why it was recalled
Siemens has confirmed that the TACR method may demonstrate reduced on-board stability which may result in imprecise and inaccurate QC and patient results.
Root cause (FDA determination)
Component design/selection
Action the firm took
Siemens sent an Urgent Medical Device Correction letter dated April 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. A recall notification letter was sent to consignees in April 2013. Customers using the affected product should load Flex reagent cartridges on board immediately prior to use and remove the Flex after 8 hours. Do not load more Flexes than the lab will use in 8 hours and do not pre-hydrate TACR Flexes. Customers were asked to complete the attached Effectiveness Check form and fax it to the Siemens Technical Solutions Center at 302-631-8467. Customers were asked to forward the communication to anyone to whom may have distributed the product. Customers with questions were instructed to contact the Siemens Technical Solutions Center or their local Siemens technial support representative. For questions regarding this recall call 800-441-9250.
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Address
- 500 Gbc Dr PO BOX 6101, Ms 514, Newark, Delaware 19702-2466
Distribution
- Distribution pattern
- Nationwide Distribution including AK, AL, AR, AZ, CA, DC, FL, GA, IA, IL, LA, MD, MI, MO, MS, NC, NH, NJ, NM, OH, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, and WI.
Timeline
- Recall initiated
- 2013-04-25
- Posted by FDA
- 2013-12-16
- Terminated
- 2014-11-12
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #117800. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.