Recalls / —
—#118055
Product
Canister: CO2 canister with GE part numbers 1407-3200-000, 1407-7004-000, or M1084850. Intended to provide general inhalation anesthesia and ventilatory support.
- FDA product code
- BSZ — Gas-Machine, Anesthesia
- Device class
- Class 2
- Medical specialty
- Anesthesiology
- 510(k) numbers
- K090892, K110213, K112722, K122445, K123125
- Affected lot / code info
- Canisters with date code of 02-2013 (Feb 2013)
Why it was recalled
GE Healthcare has recently become aware of a potential safety issue involving the reusable CO2 absorbent canister accessory used with Aespire, Aespire View, Avance, Avance CS2, Aisys, ADU, 9100, M900, M904, 9300 systems. Reusable CO2 absorbent canisters accessories may not seal properly.
Root cause (FDA determination)
Component change control
Action the firm took
GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter April 30, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Fax back form was included with the letter requesting the consignees to complete and return.
Recalling firm
- Firm
- GE Healthcare, LLC
- Address
- 3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615
Distribution
- Distribution pattern
- Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, CT, ID, MK, MI, NJ, NY, OH, TN, TX, and VA, and the countries of INDIA, AUSTRALIA, BAHRAIN, BANGLADESH, BOLIVIA, BOSNIA & HERZEGOWINA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, HONG KONG, HUNGARY, INDONESIA, IRAQ,ISRAEL, ITALY, JAPAN, REPUBLIC OF KOREA,LEBANON, MALAYSIA, MEXICO, NETHERLAND,NEW ZEALAND, OMAN, PARAGUAY, PERU., PHILIPPINES, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, THAILAND, TURKEY, UNITED ARAB EMIRATES , UNITED KINGDOM, and VIET NAM.
Timeline
- Recall initiated
- 2013-05-14
- Posted by FDA
- 2013-06-03
- Terminated
- 2014-03-12
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #118055. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.