FDA Device Recalls

Recalls /

#118581

Product

Stryker Instruments 0400-820-100 T5 Zipper Toga with Peel Away, Sterile, O; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.

FDA product code
FYAGown, Surgical
Device class
Class 2
Medical specialty
General, Plastic Surgery
510(k) numbers
K070078
Affected lot / code info
1006130, 1101034, 1006134, 1102133, 1007151, 1103208, 1007170, 1104297, 1009205, 1105360, 1010228, 1106476, 1010260, 1107569, 1011296, 1108676, 1012336, 11097861006141, 1012303, 1102110, 1104324, and 1108664.

Why it was recalled

Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Stryker sent an Urgent Urgent Recall noitifications letter dated May 7, 2013 via certified US mail and E-mail l to all affected customers, distributors and direct accounts. The letter identified the affected products, problem, risk to health, and actions to be taken. Customers were instructed to check inventory, immediately quarantine any affected product, and open each package before discarding to prevent use. Customers were directed to fill out and return a Business Reply Form to Stericycle, Phone: 888-943-2396 Fax: 888-943-4192, Email: stryker5893@stericycle.com verifying directions and confirming recalled product was destroyed. For questions call 269-389-4354 or 269-389-4518.

Recalling firm

Firm
Stryker Instruments Div. of Stryker Corporation
Address
4100 E Milham Ave, Portage, Michigan 49002-9704

Distribution

Distribution pattern
USA Nationwide Distribution

Timeline

Recall initiated
2013-05-23
Posted by FDA
2013-06-06
Terminated
2014-04-14
Status

Source: openFDA Device Recall endpoint. Recall record ID #118581. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.