Product

00597504110 Provisional AC ART SURF PROV 56/STR GRN 10, Rx, Sterile; 00597504112 Provisional AC ART SURF PROV 56/STR GRN 12, Rx, Sterile; 00597504114 Provisional AC ART SURF PROV 56/STR GRN 14, Rx, Sterile; 00597504117 Provisional AC ART SURF PROV 56/STR GRN 17, Rx, Sterile; 00597504120 Provisional AC ART SURF PROV 56/STR GRN 20, Rx, Sterile; 00597504123 Provisional AC ART SURF PROV 56/STR GRN 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

FDA product code
JWHProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
Affected lot / code info
all codes

Why it was recalled

Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d

Root cause (FDA determination)

Labeling design

Action the firm took

Beginning June 7, 2013 all distributors will be sent Urgent Device Recall notices via electronic mail, with distributors, hospital risk managers with affected inventory also receiving the notice via courier. The Notice outlined the products recalled; the reason for the recall, possible adverse events related to the recall and directed consignees to locate, quarantine and count all affected product recording the information on the Inventory Return Form acknowledging that recall directions were followed. Further questions should be directed to 1-877-946-2761. Affected product is to be returned to Zimmer Product Service Department , 1777 West Center St. Warsaw, IN 46580 or internation return at Zimmer International Logistics GmbH, Max - Immelmann Allee 12 79427 Eschbach Germany. " Communications outside of the United States will occur approximately two weeks after the United States communications.

Recalling firm

Firm
Zimmer, Inc.
Address
345 E Main St, Warsaw, Indiana 46580-2746

Distribution

Distribution pattern
Worldwide Distribution-USA (nationwide) and the country of Japan.

Timeline

Recall initiated
2013-06-07
Posted by FDA
2013-08-22
Terminated
2014-12-15
Status