FDA Device Recalls

Recalls /

#119676

Product

LPS Lower Extremity Dovetail Intercalary component. Intended for use in replacement of the mid-shaft portion of the femur.

FDA product code
JWHProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K041085
Affected lot / code info
CATALOG NO. : Part Number: 1987-19-055 and lot numbers:130896, 132131, 132133, 209461, 209466, 219843, 232972, 295965, 310189, 336843, 349283, 352878, 374125, 379089, A1TAR1, A2AGN1, A2AGS1, A2AGT1, A2AGV1, B2VAL1, B43G11, B43G1A, B43G1B, B43G1C, B43G1D, B69HV1, B69HVA, BK1BB1, BW7DX1, BW7DX1A, E3SJD1, and EN4KJ1.  In certain documentation, these alphanumeric lot numbers may appear with three zeros at the end (e.g. A1TAR1 may appear as A1TAR1000).

Why it was recalled

The LPS Lower Extremity Dovetail Intercalary component has the potential for fracture of the female component when exposed to certain physiological loads.

Root cause (FDA determination)

Device Design

Action the firm took

On 7/11/13, URGENT MEDICAL DEVICE RECALL NOTIFICATIONS were sent to consignees: Depuy Distributors, hospitals, and surgeons. End-user Hospital Customers will submit reconciliation forms to DePuy. Possible clinical implications of device failure are stated in these notifications, including poor mechanics and loss of function, pain, component dislocation, adverse tissue reaction, bone damage, potential need for revision surgery and the risks associated with this type of surgery. A patient letter will be included with the surgeon letter to assist in surgeon's notification and discussion with patients. The devices will be returned to DePuy Warsaw through the normal DePuy Returns process, to attention of Returns and marking H13-13 on the outside of the box. If a patient presents with a fractured LPS" Lower Extremity Dovetail Intercalary component with well-fixed proximal and distal stems and the surgeon determines LPS" Lower Extremity Dovetail Intercalary component is the best treatment option, DePuy plans to make the LPS" Lower Extremity Dovetail Intercalary component available through authorized approvals. Directions on how to purchase the component are provided in the Recall Communications. Any questions or concerns about the recall, please contact Kim Earle, Recall Coordinator, at 574-371-4917 (M-F; 8am-5pm EDT)

Recalling firm

Firm
DePuy Orthopaedics, Inc.
Address
700 Orthopaedic Dr, Warsaw, Indiana 46582-3994

Distribution

Distribution pattern
Worldwide Distribution-USA including the states of FL, GA, ID, KS, KY, MI, MO, NC, NV, NY, OH, OK, PA, TN, TX, VA, and WA and the countries of AUSTRALIA, CANADA, ENGLAND, GERMANY, INDIA, COUNTY CORK, IRELAND, and TURKEY.

Timeline

Recall initiated
2013-07-11
Posted by FDA
2013-07-31
Terminated
2014-07-10
Status

Source: openFDA Device Recall endpoint. Recall record ID #119676. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.