Recalls / —
—#119941
Product
Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of tacrolimus therapy in kidney and liver transplant patients.
- FDA product code
- MLM — Enzyme Immunoassay, Tracrolimus
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K060502
- Affected lot / code info
- TACR Flex Reagent cartridge, material #10444938, (DF107) lot BB4087 with expiration date 3/28/14 and its linked calibrator, material #10445012, (DC 107) lot 3BD029 with expiration date 03/01/14.
Why it was recalled
Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing.
Root cause (FDA determination)
Process control
Action the firm took
Siemens sent an Urgent Medical Device Recall letter dated July 2013 to affected customers The letter identified the affected product, problem, potential risk, and actions to be taken by customers. Customers were requested to discuss the letter with thier laboratory director to determine if additional follow up is appropriate for patient tests conducted with the affected device lots. Additionally, customers were asked to discard any remaining inventory and to complete and return the field correction effectiveness check for replacement product. They were also asked to forward the letter to anyone wha had received this affected product. For questions contact your Siemens Technical Solutions Center or your local Siemens Technical support representative.
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Address
- 500 Gbc Dr PO BOX 6101, Ms 514, Newark, Delaware 19702-2466
Distribution
- Distribution pattern
- USA Nationwide Distribution including the states of: TX, NC, GA, TN, UT, NJ, MI, VA, AR, LA, MO, FL, AZ, CA, DC, PA, NH, WI, SD, SC, NM, MD, and IN.
Timeline
- Recall initiated
- 2013-07-01
- Posted by FDA
- 2013-08-07
- Terminated
- 2014-08-26
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #119941. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.