Recalls / —
—#120690
Product
NexGen Femoral Component (CR-FLEX GSF PRECOAT SZ C-L; CR-FLEX GSF PRECOAT SZ C-R ; CR-FLEX GSF PCT SZ C-L MINUS; CR-FLEX GSF PCT SZ C-R MINUS ;CR-FLEX GSF PRECOAT SZ D-L; CR-FLEX GSF PRECOAT SZ D-R ;CR-FLEX GSF PCT SZ D-L MINUS ; CR-FLEX GSF PCT SZ D-R MINUS ;CR-FLEX GSF PRECOAT SZ E-L CR-FLEX GSF PRECOAT SZ E-R ;CR-FLEX GSF PCT SZ E-L MINUS ;CR-FLEX GSF PCT SZ E-R MINUS ;CR-FLEX GSF PRECOAT SZ F-L ;CR-FLEX GSF PRECOAT SZ F-R ;CR-FLEX GSF PCT SZ F-L MINUS ;CR-FLEX GSF PCT SZ F-R MINUS ;CR-FLEX GSF PRECOAT SZ G-L ;CR-FLEX GSK PRECOAT SZ G-R ;CR-FLEX GSF PCT SZ G-L MINUS ;CR-FLEX GSF PCT SZ G-R MINUS ; LPS-FLEX GSF OPT SZ C-L ; LPS-FLEX GSF OPT SZ C-R ;LPS-FLEX GSF OPT SZ D-L ;LPS-FLEX GSF OPT SZ D-R ;LPS-FLEX GSF OPT SZ E-L ;LPS-FLEX GSF OPT SZ E-R ;LPS-FLEX GSK OPT SZ F-L ;LPS-FLEX GSF OPT SZ F-R ;LPS-FLEX GSF OPT SZ G-L ; LPS-FLEX GSF OPT SZ G-R ;CR-FLEX PCT FEM B-L; CR-FLEX PCT FEM B-R; CR-FLEX PCT FEM C-L; CR-FLEX PCT FEM C-R; CR-FLEX PCT FEM C-L MINUS; CR-FLEX PCT FEM C-R MINUS; CR-FLEX PCT FEM D-L; CR-FLEX PCT FEM D-R; CR-FLEX PCT FEM D-L MINUS; CR-FLEX PCT FEM D-R MINUS; CR-FLEX PCT FEM E-L; CR-FLEX PCT FEM E-R; CR-FLEX PCT FEM E-L MINUS; CR-FLEX PCT FEM E-R MINUS; CR-FLEX PCT FEM F-L; CR-FLEX PCT FEM F-R; CR-FLEX PCT FEM F-L MINUS; CR-FLEX PCT FEM F-R MINUS; CR-FLEX PCT FEM G-L; CR-FLEX PCT FEM G-R; CR-FLEX PCT FEM G-L MINUS; CR-FLEX PCT FEM G-R MINUS; CR-FLEX POR FEM B-L; CR-FLEX POR FEM B-R; CR-FLEX POR FEM C-L; CR-FLEX POR FEM C-R; CR-FLEX POR FEM C-L MINUS; CR-FLEX POR FEM C-R MINUS; CR-FLEX POR FEM D-L; CR-FLEX POR FEM D-R; CR-FLEX POR FEM D-L MINUS; CR-FLEX POR FEM D-R MINUS; CR-FLEX POR FEM E-L; CR-FLEX POR FEM E-R; CR-FLEX POR FEM E-L MINUS; CR-FLEX POR FEM E-R; MINUS; CR-FLEX POR FEM F-L; CR-FLEX POR FEM F-R ;CR-FLEX POR FEM F-L MINUS; CR-FLEX POR FEM F-R MINUS; CR-FLEX POR FEM G-L; CR-FLEX POR FEM G-R; CR-FLEX POR FEM G-L MINUS CR-FLEX POR FEM G-R MINUS; CR-FLEX OPT FEM B-L; CR-FLEX OPT FEM B-R; CR-FLEX OPT FEM C-L; CR-FLEX OPT FEM C-R; CR-FLEX OPT FEM C-L MINUS; CR-FLEX OPT FEM C-R MINUS; CR-FLEX OPT FEM D-L; CR-FLEX OPT FEM D-R; CR-FLEX OPT FEM D-L MINUS; CR-FLEX OPT FEM D-R MINUS; CR-FLEX OPT FEM E-L; CR-FLEX OPT FEM E-R; CR-FLEX OPT FEM E-L MINUS; CR-FLEX OPT FEM E-R MINUS; CR-FLEX OPT FEM F-L; CR-FLEX OPT FEM F-R; CR-FLEX OPT FEM F-L MINUS; CR-FLEX OPT FEM F-R MINUS; CR-FLEX OPT FEM G-L; CR-FLEX OPT FEM G-R; CR-FLEX OPT FEM G-L MINUS; CR-FLEX OPT FEM G-R MINUS; LPS-FLEX CO-NID FEM A-L ; LPS-FLEX CO-NID FEM A-R ; LPS-FLEX RECOAT FEMORAL A-L; LPS-FLEX PRECOAT FEMORAL A-R; LPS-FLEX CO-NID FEM B-L ; LPS-FLEX CO-NID FEM B-R ; LPS-FLEX PRECOAT FEMORAL B-L; LPS-FLEX PRECOAT FEMORL B-R; LPS-FLEX CO-NID FEM C-L ; LPS-FLEX CO-NID FEM C-R ; LPS-FLEX PRECOAT FEMORAL C-L; LPS-FLEX PRECOAT FEMORAL C-R; LPS-FLEX CO-NID FEM D-L ; LPS-FLEX CO-NID FEM D-R ; LPS-FLEX PRECOAT FEMORAL D-L; LPS-FLEX PRECOAT FEMORAL D-R; LPS-FLEX CO-NID FEM E-L ; LPS-FLEX CO-NID FEM E-R ; LPS-FLEX PRECOAT FEMORAL E-L; LPS-FLEX PRECOAT FEMORAL E-R; LPS-FLEX CO-NID FEM F-L ; LPS-FLEX CO-NID FEM F-R ; LPS-FLEX PRECOAT FEMORAL F-L; LPS-FLEX PRECOAT FEMORAL F-R; LPS-FLEX CO-NID FEM G-L ; LPS-FLEX CO-NID FEM G-R ; LPS-FLEX PRECOAT FEMORAL G-L; LPS-FLEX PRECOAT FEMORAL G-R; LPS-FLEX PRECOAT FEMORAL H-L; LPS-FLEX PRECOAT FEMORAL H-R ; LPS-FLEX POROUS FEMORAL B-L ; LPS-FLEX POROUS FEMORAL B-R ; LPS-FLEX POROUS FEMORAL C-L ; LPS-FLEX POROUS FEMORAL C-R LPS-FLEX POROUS FEMORAL D-L ; LPS-FLEX POROUS FEMORAL D-R ; LPS-FLEX POROUS FEMORAL E-L ; LPS-FLEX POROUS FEMORAL E-R ; LPS-FLEX POROUS FEMORAL F-L ; LPS-FLEX POROUS FEMORAL F-R ;LPS-FLEX POROUS FEMORAL G-L ; LPS-FLEX POROUS FEMORAL G-R ; LPS-FLEX POROUS FEMORAL H-L ; LPS-FLEX POROUS FEMORAL H-R ; LPS-FLEX OPTION FEMORAL A-L; LPS-FLEX OPTION FEMORAL A-R; LPS-FLEX OPTION FEMORAL B-L; LPS-FLEX OPTION FEMORAL B-R; LPS-FLEX OPTION FEMORAL B-L ; LPS-FLEX OPTION FEMORAL B-R ; LPS-FLEX OPTION FEMORAL C-L; LPS-FLEX OPTION FEMORAL C-R; LPS-FLEX OPTION FEMORAL C-L; LPS-FLEX OPTION FEMORAL C-R; LPS-FLEX OPTI
- FDA product code
- JWH — Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K060370
- Affected lot / code info
- Part Number Range: 005750, 5764, 5950, 5952, 5956, 5960, 5962, 5964, 5966, 5970, 5972, 5990, 5992, 5994, and 5996-011/018-01/92.
Why it was recalled
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
Root cause (FDA determination)
Package design/selection
Action the firm took
August 30, 2013 Urgent Recall Notices were sent to Distributors, hospital risk managers and surgeons identifying the issue and directing consignees to make all affected personnel aware of the contents of the notification. The notice identified the products involved, the issue, possible adverse events associated with the issue and recommended that implants found with LDPE bag adhesion not use the product and return it to Zimmer sales representatives. Questions or concerns should be addressed to Zimmer customer call center at 1-877-946-2761 between the hours of 8 a.m. and 5 p.m. EST.
Recalling firm
- Firm
- Zimmer, Inc.
- Address
- 345 E Main St, Warsaw, Indiana 46580-2746
Distribution
- Distribution pattern
- Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, India, Israel, Jamaica, Japan, Korea, Malaysia, Mexico, New Zealand, Nicaragua, Panama, Peru, Singapore, Taiwan, Thailand, and Venezula.
Timeline
- Recall initiated
- 2013-08-30
- Posted by FDA
- 2013-10-25
- Terminated
- 2014-09-25
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #120690. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.