FDA Device Recalls

Recalls /

#121364

Product

TWINFIX Ultra PLLA/HA 6.5 mm with 3 ULTRABRAID Suture; Part Number: 72202610 Biodegradable suture anchor

FDA product code
MAIFastener, Fixation, Biodegradable, Soft Tissue
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K093844
Affected lot / code info
50325166 50325167 50327043 50336737 50336772 50339647 50339842 50341976 50385698 50397465 50412290 50417528 50426452 50427818 50429624 50429874 50431563 50453525

Why it was recalled

Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.

Root cause (FDA determination)

Packaging

Action the firm took

Smith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include their contact information in the spaces provided. Customers were instructed to contact Smith & Nephew's Demo Service Departmentat 800-343-5717 for a Return Authorization (RA#), and provide the information response form. Replacement product would be provided. For questions regarding this recall call 978-749-1073.

Recalling firm

Firm
Smith & Nephew, Inc. Endoscopy Division
Address
150 Minuteman Drive, Andover, Massachusetts 01810-1031

Distribution

Distribution pattern
Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.

Timeline

Recall initiated
2013-08-06
Posted by FDA
2013-09-17
Terminated
2016-04-20
Status