FDA Device Recalls

Recalls /

#121601

Product

Siemens RAPIDLab 1265 system blood gas analyzer Product Usage: RAPIDLab 1200 Systems: The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1240 pH, pC02, p02 RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1260 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb

FDA product code
CHLElectrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Device class
Class 2
Medical specialty
Clinical Chemistry
510(k) numbers
K073537
Affected lot / code info
All Serial Numbers

Why it was recalled

RAPIDPoint 400 Series, RAPIDPoint 500, or RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample

Root cause (FDA determination)

Software Manufacturing/Software Deployment

Action the firm took

Siemens Healthcare sent an Urgent Field Safety Notification letter dated September 2013 via Federal Express, to affected Siemens domestic customers. Effected Siemens customers outside the US was provided a copy of the Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. Previously Siemens Healthcare had issued a Customer Notification on August 23, 2012. The letter identified the affected product, issue, information to mitigate the issue and actions to be taken. The firm will issue an upgrades software to users. For questions contact your local Customer Care Center - Technical Support Representative at 1-877-229-3711.

Recalling firm

Firm
Siemens Healthcare Diagnostics Inc
Address
2 Edgewater Dr, Norwood, Massachusetts 02062-4637

Distribution

Distribution pattern
Worldwide Distribution - USA Nationwide and in the countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Brazil, Brunei, Bulgaria, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Fren.Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Mayotte, Netherlands, New Caledonia, New Zealand, Norway, P.R. China, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, U.A.E., United Kingdom, Vatikancity, Venezuela, Vietnam,

Timeline

Recall initiated
2012-09-17
Posted by FDA
2013-10-29
Terminated
2015-04-01
Status

Source: openFDA Device Recall endpoint. Recall record ID #121601. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.