Product
Smiths Medical, Deltec: REF 21-8011-24 Port-A-Cath Dual-Lumen. Titanium Venous Access System. REF 21-8052-24 & REF 21-8053-24 , Port-A-Cath II Dual-Lumen, Polysulfone /Titanium Venous Access System. Sterile EO, Rx Only. PORT-A-CATH and PORT-A-CATH II implantable venous and arterial access systems are designed to permit repeated access to the vascular system for the parenteral delivery of medications, fluids, and nutritional solutions and for the sampling of venous blood. Introducer sets are indicated for the introduction of catheters into the subclavian vein.
- FDA product code
- LJT — Port & Catheter, Implanted, Subcutaneous, Intravascular
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K932095, K932841, K955407
- Affected lot / code info
- 21-8011-24: 1971567 1979805 1984078 1997124 2006124 2038001 2090898 2151160 2196530 2216256 2249695 21-8052-24: 1971572 2012472 2022904 2067947 2080704 2146472 2168988 2196532 2207404 2237458 2249696 2269178 2282158 21-8053-24: 2168989 2249697
Why it was recalled
Smiths Medical is conducting a voluntary recall of PORT-A-CATH and PORT-A-CATH II Implantable Venous and Arterial Access Systems and Introducer Sets. Smiths Medical has become aware of a small number of customer reports regarding the introducer sheath and the introducer opening being too small for the width of the catheter with the use of these Sets. Smiths Medical has received no reports
Root cause (FDA determination)
Mixed-up of materials/components
Action the firm took
Smiths Medical ASD, Inc. sent an Urgent Field Safety Notice dated September 6, 2013, to all affected customers. The letter described the problem and the product involved in the recall. The letter included the "Advice on Action to be Taken by the User" which instructed consignees to inspect and remove affected devices and to complete and return the "Urgent Field Safety Notice Confirmation Form. The letter also included information on the "Transmission of this Urgent Field safety Notice" which included information where customers should report issues related to the product to Smiths Medical's Global Complaint Department at 1-866-216-8806 or globalcomplaints@smiths-medical.com. For questions on the information provided they should contact 800-258-5361 (Smiths Medical's Customer Service Department). For questions regarding this recall call 1-800-332-0178.
Recalling firm
- Firm
- Smiths Medical ASD, Inc.
- Address
- 1265 Grey Fox Rd, Saint Paul, Minnesota 55112-6929
Distribution
- Distribution pattern
- Worldwide Distribution - USA including AZ, AR, CA, GA, IL, IN, KY, MD, MA, MI, NV, NJ, NY, NC, OH, OR, PA, RI, SC, TN, TX, VA, WA, WV, WY. Internationally to DENMARK, BELGIUM CANADA.
Timeline
- Recall initiated
- 2013-09-10
- Posted by FDA
- 2013-09-23
- Terminated
- 2014-10-21
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #121807. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.