FDA Device Recalls

Recalls /

#122414

Product

*** 1) Product labeled in part: List No. 14220-28; PLUM Y-TYPE BLOOD SET; 200 Micron Filter, CLAVE Port, CLAVE Y-Site, Non-Vented, 110 Inch, Non-DEHP; Product Usage: The CLAVE of the secondary port is used as an easy secondary access that allows concurrent and piggyback mode in infusion. It is normally used with a syringe or a secondary IV infusion set.

FDA product code
FPASet, Administration, Intravascular
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K101677
Affected lot / code info
***** CANADA DISTRIBUTION *****   *** 1) List Number 14220-28-01

Why it was recalled

Intravenous fluid may leak at the CLAVE secondary port of Hospira PlumSet IV administration sets.

Root cause (FDA determination)

Device Design

Action the firm took

Hospira sent an URGENT MEDICAL DEVICE CORECTION letters dated September 9, 2013 to customers. The letter identified the affected product, problem and actions to be taken. the letter included additional instructions on the proper handling/use of the CLAVE devices. The letters request that all incidents of separation or breakage of CLAVE from the secondary port of the cassette be reported to Hospira by calling 1-800-441-410 (Monday - Friday, 8:00 AM - 5:00 PM CST) or by e-mail at ProductComplaintPP@hospira.com. Customers are further requested to complete and return a response form included wih the URGENT MEDICAL DEVICE CORRECTION letter and to notify any customers of this advisory. Hospira will also be supplementing the instructions for CLAVE use as an interim measure until a new method for bonding the CLAVE to the secondary port of the Plum cassettes can be devised and implemented.

Recalling firm

Firm
Hospira Inc.
Address
275 N Field Dr, Lake Forest, Illinois 60045-2579

Distribution

Distribution pattern
US (Nationwide) Distribution

Timeline

Recall initiated
2013-09-09
Terminated
2016-12-06
Status