Product
Hospira RECEPTAL CANISTERS: 1L(1000mL), 1.5L(1500mL), 2L(2000mL) Canister, ATS, 2000, with Valve 1) List Numbers 43449-01 and 43449-11; LIST NO. 43449 RECEPTAL CANISTER; CANISTER FOR REUSE, THROW AWAY LINER ONLY; DO NOT FILL 900 ABOVE LINE 2) List Numbers 43423-01 and 43423-21; LIST NO. 43423; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; 1400. 3) List Numbers 43445-01 and 43445-11; LIST NO. 43445; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; DO NOT FILL 1900 ABOVE THIS LINE. 4) List Number 43496-01.
- FDA product code
- GCX — Apparatus, Suction, Operating-Room, Wall Vacuum Powered
- Device class
- Class 2
- Medical specialty
- General Hospital
- Affected lot / code info
- 43423-04-01 43423-04-11 43423-04-21 43445-04-01 43445-04-11 43449-04-01 43449-04-11 43496-04-01
Why it was recalled
Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.
Root cause (FDA determination)
Process control
Action the firm took
URGENT MEDICAL DEVICE RECALL letters dated September 10, 2013 were distributed to direct accounts notifying them of the recall. The letter requests that customers to ensure that they only use liners and canisters of corresponding sizes (e.g. a 1000mL liner with a 1000mL canister) and to only use Hospira components. The letter further advises customers of an ongoing recall of the 1000mL Receptal systems. Customers were requested to complete and return the attached reply form and to notify other healthcare providers within their organization.Customers who further distributed the Receptal systems were requested to forward the letter to their customers and to request that the contact Stericycle at 1-888-943-5175 (Monday - Friday, 8:00 AM - 5:00 PM, CST) to obtain a reply form.
Recalling firm
- Firm
- Hospira Inc.
- Address
- 275 N Field Dr, Lake Forest, Illinois 60045-2579
Distribution
- Distribution pattern
- *** US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY), District of Columbia, Puerto Rico; *** FOREIGN: Australia, Canada, Costa Rica, Singapore, South Africa
Timeline
- Recall initiated
- 2013-09-10
- Terminated
- 2016-12-15
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #122416. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.