Recalls / —
—#122630
Product
THUNDER Steerable Guidewire; Model: LVTNDR190S Medtronic guide wires are steerable guide wires that are used for the introduction and placement of diagnostics or interventional devices in the coronary and peripheral vasculature, and may be used to reach and cross a target lesion. Medtronic guide wires are not intended for cerebral vasculature. Medtronic steerable exchange guide wires are used to facilitate the substitution of one diagnostic or interventional device for another.
- FDA product code
- LIT — Catheter, Angioplasty, Peripheral, Transluminal
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K970466
- Affected lot / code info
- G13A00116 G13A00936 GV260 GV846
Why it was recalled
Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
The firm, Medtronic, sent a "URGENT MEDICAL DEVICE RECALL" letter dated October 21, 2013 (in the US via UPS overnight service, 1-day delivery) to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately remove and quarantine all potentially affected guidewires that remain in inventory; return the affected product to Medtronic; if replacement product is needed, your Medtronic representative can assist you with identifying suitable replacement product, and to complete and return the Customer Confirmation Certificate confirming receipt of the notification via fax to Medtronic at 651-367-0612 to the attention of Customer Focused Quality (100% customer confirmations will be required). Should have any questions, please contact your Medtronic Representative or call 763-526-2513. Notifications to customers concerning the additional affected product began December 19, 2013 (in the US via UPS overnight service, 1-day delivery). Customers are requested to return a Confirmation Certificate confirming their receipt of the notification. 100% customer confirmations will be required from the additional consignees.
Recalling firm
- Firm
- Medtronic Vascular
- Address
- 35-37A Cherry Hill Drive, Danvers, Massachusetts 01923-2565
Distribution
- Distribution pattern
- Worldwide distribution: US (Nationwide) including states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV; and countries of: Albania, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guyana, Honduras, Hong Kong,Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malaysia, Mexico, Montenegro, Nepal, Netherlands, New Zealand, Nepal, Nicaragua, Norway, Oman, Poland, Portugal Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, United Kingdom, Uruguay, and Yemen.
Timeline
- Recall initiated
- 2013-10-21
- Posted by FDA
- 2013-11-15
- Terminated
- 2014-12-15
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #122630. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.